Levelflex FMP56, FMP57 FOUNDATION Fieldbus
Mounting
Hauser
31
Installation in nozzles ≥ DN300
If installation in ≥ 300mm/12" nozzles is unavoidable, installation must be carried out in
accordance with the following sketch.
1
2
3
4
A0014199
1
Lower edge of the nozzle
2
Approx. flush with the lower edge of the nozzle (± 50 mm/2")
3
Plate
4
Pipe 150 to 180 mm (6 to 7 inch)
Nozzle diameter
Plate diameter
300 mm (12")
280 mm (11")
≥ 400 mm (16")
≥ 350 mm (14")
Содержание Levelflex FMP56