Operating menu
Levelflex FMP56, FMP57 FOUNDATION Fieldbus
166
Hauser
"Probe settings" submenu
The
Probe settings
submenu helps to ensure that the end of probe signal within the
envelope curve is correctly assigned by the evaluation algorithm. The assignment is correct
if the length of probe indicated by the device matches the acutal lenght of the probe. The
automatic probe length correction can only be performed if the probe is installed in the
vessel and is completely uncovered (no medium). For partially filled vessels and if the
probe length is known, select
Confirm probe length (→ 167)
=
Manual input
in order
to enter the value manually.
If a mapping (interference echo suppression) has been recorded after shortening the
probe, it is no longer possible to perform an automatic probe length correction. In this
case there are two options:
• Delete the map using the
Record map
parameter (→ 142) before performing the
automatic probe length correction. After the probe length correction, a new map
can be recorded using the
Record map
• Alternative: Select
Confirm probe length (→ 167)
Manual input
and enter
the probe length manually into the
Present probe length
An automatic probe length correction is only possible after the correct option has
been selected in the
Probe grounded
Navigation
Setup → Advanced setup → Probe settings
Probe grounded
Navigation
Setup → Advanced setup → Probe settings → Probe grounded
Prerequisite
Operating mode
=
Level
Description
Specify whether the probe is grounded.
Selection
• No
• Yes
Factory setting
No
Present probe length
Navigation
Setup → Advanced setup → Probe settings → Pres. length
Description
• In most cases:
Displays the length of the probe according to the currently measured end-of-probe
signal.
• For
Confirm probe length (→ 167)
=
Manual input
:
Enter actual length of probe.
User entry
0 to 200 m
Factory setting
4 m