Levelflex FMP56, FMP57 FOUNDATION Fieldbus
Operating menu
Hauser
153
"Linearization" submenu
0%
100%
L
L
L
L
L
L
L
L2
L1
L‘1L‘2
L‘
L‘
L‘
L‘
L‘
L‘
L‘
M
M
M
M
M
H
h
h
M
0%
100%
100%
100%
0%
0%
100%
0%
0%
0%
100%
100%
B
A
C
D
E
F
G
H
2
1
L
I
L‘
I‘
d
d
A0016084
39
Linearization: Transformation of the level and (if relevant) the interface height into a volume or weight;
the transformation is dependent on the shape of the vessel.
1
Selection of linearization type and unit
2
Configuration of the linearization
A
Linearization type (→ 156) = None
B
Linearization type (→ 156) = Linear
C
Linearization type (→ 156) = Table
D
Linearization type (→ 156) = Pyramid bottom
E
Linearization type (→ 156) = Conical bottom
F
Linearization type (→ 156) = Angled bottom
G
Linearization type (→ 156) = Horizontal cylinder
H
Linearization type (→ 156) = Sphere
I
For "Operating mode" = "Interface" or "Interface with capacitance": Interface before linearization (measured in
distance units)
I'
For "Operating mode" = "Interface" or "Interface with capacitance": Interface after linearization (corresponds to
volume or weight)
L
Level before linearization (measured in distance units)
L'
Level linearized (→ 158) (corresponds to volume or weight)
M
Maximum value (→ 159)
d
h