Deltapilot S FMB70
Electrical connection
Hauser
11
• Guide the cable through the gland. Preferably use a twisted, shielded two-wire cable.
• Connect the device as indicated in the diagram.
• Screw down the housing cover.
• Switch on the supply voltage.
4... 20mA
Test
Test
Test
4... 20mA
Test
1
8
6
7
2
3
4
5
A0019989
1
Housing
2
Minimum supply voltage = 10.5 V DC, jumper is set as illustrated in the diagram.
2
Minimum supply voltage = 11.5 V DC, jumper is set to the "Test" position.
3
4 to 20 mA
4
Devices with integrated overvoltage protection are labeled "OVP" (overvoltage protection) here.
5
External ground terminal
6
4 to 20 mA test signal between positive and test terminal
7
Internal ground terminal
8
Jumper for 4 to 20 mA test signal
Содержание FMB70