Electrical connection
Deltapilot S FMB70
10
Hauser
5.2.3
Closing the housing covers
NOTICE
Devices with EPDM cover seal - leaking transmitter!
Mineral-, animal- or plant-based lubricants cause the EPDM cover seal to swell and the
transmitter to leak as a result.
‣
It is not necessary to grease the thread due to the coating applied to the thread at the
factory.
NOTICE
The housing cover can no longer be closed.
Damaged thread!
‣
When closing the housing covers make sure that the threads on the covers and the housing
are free from dirt, such as sand. If you encounter resistance when closing the covers, then
check the threads again for dirt.
Closing the covers on the hygienic stainless steel housing (T17)
The covers for the terminal compartment and electronics compartment are hooked into the
housing and closed with a screw in each case. These screws must be tightened finger-tight (2
Nm (1.48 lbf ft)) to the stop to ensure that the covers are securely seated and leak-tight.
6
Electrical connection
6.1
Connecting the device
L
WARNING
Risk of electric shock!
If the operating voltage is > 35 VDC: Dangerous contact voltage at terminals.
‣
In a wet environment, do not open the cover if voltage is present.
L
WARNING
An incorrect connection compromises electrical safety!
‣
• Risk of electric shock and/or explosion! Switch off the supply voltage before connecting the
device.
• When using the measuring device in hazardous areas, installation must also comply with
the applicable national standards and regulations and the Safety Instructions or Installation
or Control Drawings.
• Devices with integrated overvoltage protection must be grounded.
• Protective circuits against reverse polarity, HF influences and overvoltage peaks are
integrated.
• The supply voltage must match the power supply on the nameplate, see Operating
Instructions.
• Switch off the supply voltage before connecting.
• Remove the housing cover of the terminal compartment.