Deltapilot S FMB70
Mounting
Hauser
9
Level measurement
• Always install the device below the lowest measuring point.
• Do not install the device at the following positions:
• In the filling curtain
• In the tank outlet
• In the suction area of a pump
• At a point in the tank which could be affected by pressure pulses from the agitator
• The adjustment and functional test can be carried out more easily if you mount the device
downstream from a shutoff device.
• The Deltapilot S must also be insulated in the case of media that can harden when cold.
Pressure measurement in gases
Mount the Deltapilot S with shutoff device above the tapping point so that any condensate
can flow into the process.
Pressure measurement in vapors
• Mount the Deltapilot S with the water pocket pipe above the tapping point.
• Fill the water pocket pipe with liquid before commissioning. The water pocket pipe reduces
the temperature to almost ambient temperature.
Pressure measurement in liquids
Mount the Deltapilot S with the shutoff device below or at the same level as the tapping point.
5.2.2
Turning the housing
The housing can be rotated by up to 380° by loosening the setscrew.
T14
T15
T17
3
2
A0019996
1. T14 and T15 housing: loosen the setscrew with a 2 mm (0.08 in) Allen key. T17
housing: loosen the setscrew with a 3 mm (0.12 in) Allen key.
2. Rotate the housing (max. up to 380°).
3. Retighten the setscrew with 1 mm (0.74 lbf ft)1 Nm (0.74 lbf ft).