Maintenance
ASP Station 2000 RPS20B
48
Hauser
Cleaning other parts conveying medium
4
4
3
3
1
1
2
2
5
A0023667
1. Internal intake hose (item 2): unscrew at the dosing system (item 1) and hose gland.
Rinse with water or soap.
2. Tap (item 3): pull out towards the front and remove the cover. Rinse with water or soap.
3. Distribution pans (item 4): pull forwards out of the frame. Rinse with water or soap.
4. Remove the bottle trays (item 5). Clean the bottles and bottle trays with water or soap.
11.2.3 Sample compartment
The sample compartment has a continuous inner plastic lining. Once the bottle trays,
distribution pan and tap have been removed, the entire sample compartment can therefore be
cleaned with a water hose.
11.2.4 Fan, liquefier
The fan and the liquefier must be purged with compressed air at regular intervals, depending
on the ambient conditions (e.g. very dusty environments).
Содержание ASP Station 2000 RPS20B
Страница 2: ......
Страница 58: ......
Страница 59: ......
Страница 60: ...www addresses endress com 71431888 71431888 ...