Diagnostics and troubleshooting
ASP Station 2000 RPS20B
44
Hauser
10
Diagnostics and troubleshooting
10.1 Troubleshooting instructions
Always start troubleshooting with the following checklists if faults occur after commissioning
or during operation. The routine takes you directly to the cause of the problem and the
appropriate remedial measures.
10.2 Process error messages
Message
Cause
Test or remedial measures
ERROR: RAM
New program transmitted
Acknowledge the message
ERROR: Clock
Electronics error
Repair by Service
ERROR: EEPROM
New program transmitted
EEPROM defective
Acknowledge the message
Repair by Service
ERROR: Conductivity 1
Contact between conductivity electrodes
due to water/dirt
Clean the dosing system
ERROR: Conductivity 2
ERROR: Air manager zero
point
The air manager cannot find the zero
position
Replace the air manager; if necessary have
repaired by Service
ERROR: Tap zero point
Tap defective or jammed
Check the tap; if necessary replace the
distribution system or have repaired by
Service
ERROR: Distributor missing
Distributor not connected to controller
Check the connector of the distributor; if
necessary, have repaired by Service
ERROR: 4-20mA < 3mA
Analog signal transmitter defective, no
analog signal connected, cable open circuit
Check signal current, line and signal
transmitter
ERROR: Temp. sample
compartment
Temperature in sample compartment >
60°C
Temperature sensor defective
Check installation conditions
Repair by Service
ERROR: Temp. too high
Temperature in upper compartment >
70°C
Temperature sensor defective
ERROR: Temp. upper
compartment
Temperature in upper compartment >
90°C
Temperature sensor defective
ERROR: Tap turned
Tap turned manually
Check tap
ERROR: Change distributor
Incorrect distribution selected in operation Check distributor and change as necessary