Operation
ASP Station 2000 RPS20B
36
Hauser
Sampling mode
A0023638-EN
16
Menu structure
Sampling can be time-paced, volume-paced or flow-paced (option) or can be triggered by an event.
Time
Sampling is triggered at configurable intervals.
Quantity
Sampling is triggered after a certain measured flow volume.
A volume signal must be connected to the analog input or one of the
digital inputs of the sampler for this function. The digital input must be
configured with the QUANTITY function for this purpose.
Ext. sign.
Sampling is triggered by an external signal.
One of the digital inputs must be assigned and configured with the
SAMPLE function.
Flow
Sampling is triggered at configurable intervals. The sample volume taken is
proportional to a flow currently measured.
The ASP Station 2000 must be fitted with a dfp dosing system ("twiddle
principle") for this function. In addition, a volume signal must be
connected to the analog input of the sampler.
Shots:
Number of samples per sample cycle.