5.6.  Calibration and Verification 

5.6.1.  Factory calibration 

Every EM TEST generator is entirely checked and calibrated as per international standard regulations before 
delivery. A calibration certificate is issued and delivered along with a list of the equipment used for the calibration 
proving the traceability of the measuring equipment. All auxiliary equipment and accessories are checked to our 
internal manufacturer guidelines. 

The calibration certificate and the certificate of compliance (if available) show the date of calibration.  

The AMETEK CTS equipment are calibrated in the factory and marked with a 
calibration mark. The used measuring instruments are traceable to the Swiss Federal 
Office of Metrology. 
The calibration date is marked. The validity of the calibration is to the responsibility 
of the user’s quality system. Neither the certificate of calibration nor the 
corresponding label mark any due date for re-calibration. 

 

Example: Calibration mark 

5.6.2.  Guideline to determine the calibration period of AMETEK CTS instrumentation 

Our International Service Departments and our QA Manager are frequently asked about the calibration interval of 
AMETEK CTS equipment. 

AMETEK CTS 

doesn’t know each customer’s Quality Assurance Policy nor do we know how often the equipment 

is used and what kind of tests are performed during the life cycle of a test equipment. Only the customer knows 
all the details and therefore the customer needs to specify the calibration interval for his test equipment. 

In reply to all these questions we like to approach this issue as follows: 
AMETEK CTS make use of a solid state semiconductor switch technique to generate high voltage transients. A 
precious advantage of this technique is the absolute lack of periodical maintenance effort. In consequence, 
thereof a useful calibration period must be defined based on two criteria: 

-  The first one is the 

customer’s Quality Assurance Policy. Any existent internal regulation must be applied at 

highest priority. In the absence of such internal regulation the utilization rate of the test equipment must be 
taken into consideration. 

 
-   Based on the experience and observation collected over the years 

AMETEK CTS recommend a calibration 

interval of 1 year

 for frequently used equipment. A 2-years calibration interval is considered sufficient for 

rarely used test generators in order to assure proper performance and compliance to the standard 
specifications. 

5.6.3.  Calibration of Accessories made by passive components only: 

Passive components do not change their technical specification during storage. Consequently, the measured 
values and the plots stay valid throughout the storage time. The date of shipment shall be considered as the date 
of calibration. 

5.6.4.  Periodic In-house verification 

Please refer to the corresponding standard before carrying out a calibration or verification. The standard 
describes the procedure, the tolerances and the necessary auxiliary means. Suitable calibration adapters are 
needed. To compare the verification results, AMETEK CTS suggests to refer to the wave shape and values of the 
original calibration certificate. 

All calibrations and verifications are always done without mains supply voltage connected to the coupling network 
input. 

 
 
 

Содержание VDS 200Q10

Страница 1: ...pply of a vehicle and complex power supply distortions in the power range up to 600W Many different waveforms are integrated as standard such as pulse 2b required by ISO 7637 and the starting profile...

Страница 2: ...r Operation V 1 00 2 35 AMETEK CTS GmbH Sternenhofstrasse 15 4153 Reinach BL1 Switzerland Phone 41 61 717 91 91 Fax 41 61 717 91 99 URL http www emtest com Copyright 2021 AMETEK CTS GmbH All right res...

Страница 3: ...3 4 Pulse 4 GM 9105 P 19 3 3 3 5 DC source 20 3 4 Service 21 3 5 Setup 22 3 6 Source Settings 23 4 Technical Data 27 4 1 Test level 27 4 2 Trigger 27 4 3 Input output 27 4 4 Interfaces 27 4 5 General...

Страница 4: ...Q10 120 60 to 60 V Imax 10 A Not supported 180 kHz VDS 200Q10 230 60 to 60 V Imax 10 A Not supported 180 kHz VDS 200Q10 1 100 60 to 60 V Imax 10 A Not supported 180 kHz VDS 200Q10 1 120 60 to 60 V Ima...

Страница 5: ...e preselected parameters The yellow button is illuminated and indicates the Test ON status After Test OFF or when no test is started the output voltage and the current will be set to zero 4 Knob Inc D...

Страница 6: ...und connection 10 Mains input 4 Cooling 11 IEEE 488 1 5 FRAME BUS 12 USB port 6 TRIGGER IN 13 CN interface 7 ANALOG IN 14 FAIL1 FAIL2 1 TEST SUPPLY OUT Output to DUT 2 Sense Sense lines to compensate...

Страница 7: ...to USB standard Therefore the user must set the same Baudrate in the device and control software Interface is supported for iso control but not for autowave control either IEEE or FRAME BUS 13 CN inte...

Страница 8: ...battery can discharge through the source This can result in a dangerous condition not only for the four quadrant source but also because you ve effectively short circuited the battery when setting th...

Страница 9: ...ng and start the test immediately The operator can first change the parameters and than start the test When leaving the test mode DC source the output voltage is automatically reset to the nominal sup...

Страница 10: ...e Drop Simulator V 1 00 00 Int V1 00 00 Waveforms SWN 001234 The serial number and the version number SWN are used for traceability reasons These numbers are listed in the factory test reports and cal...

Страница 11: ...e digit to be changed fast or slow change Pressing of the ESC button will bring the user back to the previous level from where the test can be restarted with new parameters Page 6 Start ISO Pulse 4 Af...

Страница 12: ...F7 Page 3 STANDARDS F1 ISO 7637 F2 ISO 16750 2 or WD 03 2000 2 F1 F2 F3 F4 F5 F6 F7 Pages 4 ISO 7637 ISO 16750 2 WD 03 2000 2 F1 Pulse 4 F1 Short voltage drop F2 Pulse 2b F2 Slow decrease increase F3...

Страница 13: ...Va1 60 0V 0 0 V Vb Va2 60 0V Parameters Vb 0 0V 60 0V Va1 60 0V 60 0V Va2 60 0V 60 0V t1 0 1s 99 9s t7 5ms 999ms t8 5ms 999ms t9 0 1s 99 9s t11 5ms 999ms Va 0 0V 60 0V I 1A Imax tri Auto Manual 3 3 1...

Страница 14: ...Va2 60 0V Parameters Vb 0 0V 60 0V Va1 60 0V 60 0V Va2 60 0V 60 0V t1 0 1s 99 9s t2 0 1s 99 9s t3 0 1s 99 9s int 0 1s 99 9s n 1 30 000 endl tri Auto Manual I 1A Imax 3 3 2 2 Slow decrease increase Th...

Страница 15: ...0V 60 0V Va2 59 9V 60 0V Ve 60 0V 60 0V dVa 60 0V 60 0V t1 0 1s 99 9s t2 0 1s 99 9s t3 0 1s 99 9s n 1 30 000 endl tri Auto Manual I 1A Imax 3 3 2 4 Pulse Starting profile This test simulates a motor s...

Страница 16: ...0V 0 1 60V f1 0 001Hz 180 000kHz f2 0 001Hz 180 000kHz f3 0 001Hz 180 000kHz t1 0 1s 999 9s t2 0 1s 999 9s int 0s 999s n 1 30 000 endl tri Auto Manual I 1A Imax 3 3 2 6 Overvoltage Vmax This test sim...

Страница 17: ...Sine wave Limitations Vb Vp 60 0V Vb Vp 0 0V Parameters Vb 0 0V 60 0V Vp 0 1V 60 0V t1 0 1s 99 9s f1 0 001kHz 180 00kHz f 100Hz 1 2Hz t2 1 0s 999 9s n 1 30 000 endl I 1 0A Imax Be sure to select the...

Страница 18: ...lifier can be driven by an external control voltage 10 to 10V e g from an external waveform generator For this purpose the operator must select VDS Extern by pressing the related function key The Exte...

Страница 19: ...tion to the ISO pulse 4 a 5Hz ripple of 1Vp p is superimposed during t9 to Va tr tf 10ms Input restrictions 0 0 V Vb Va1 60 0V 0 0 V Vb Va2 60 0V Parameters Vb 0 0V 60 0V Va1 60 0V 60 0V Va2 60 0V 60...

Страница 20: ...0 0V 60 0V I 1A 10A Vb I 13 5 5 F1 F2 F3 F4 F5 F6 F7 Page 5 Start DC POWER SUPPLY Vb 13 5 V I 7 0A I 5 8 A Stop ChangeZoom U 13 4 V F1 F2 F3 F4 F5 F6 F7 After Start the actual voltage and current valu...

Страница 21: ...st The operator can initialize a self test procedure to check the operation of the instrument The software will clearly explain the selftest procedure Set up The operator can change the simulator sett...

Страница 22: ...OFF Additionally the AUTO OFF function can be programmed to switch off the backlighting after a defined time when the equipment has not been in operation 1 30 min F3 Interfaces This menu will help th...

Страница 23: ...o inrush current above what is set The current limit circuitry will start immediately F7 Frequency Compensation The VDS 200Q10 x support three frequency ranges Std Standard This is the normal operatin...

Страница 24: ...AMETEK CTS VDS 200Q10 Series Manual for Operation V 1 00 24 35...

Страница 25: ...ged if the voltage is set to zero Likewise when generating a sine wave the full power of the amplifier can be applied to a capacitive load The current limit of the VDS 200Q10 x applies also to AC oper...

Страница 26: ...ture shutdown Several redundant temperature monitoring functions control the fans in multiple steps A final temperature shutdown is also implemented to prevent damage due to overheating but should gen...

Страница 27: ...thin 25 s Current limiter Ilimit Ilimit 0 Imax Voltage ripple Ur Pulses Ur 0 2 V peak to peak as per test routines 4 2 Trigger Automatic Auto release with preselected parameters Manual Manual release...

Страница 28: ...tage protection and input current Device Mains Phases Supply Voltage Ueff Primary Current Ieff Fuse VDS 200Q10 x 100 1 100 V 10 10 3 A 16 AT 115V 5 x 20 mm VDS 200Q10 x 120 1 120 V 10 8 6 A 10 AT 115V...

Страница 29: ...nal regulations The generators of the series 200 UCS LD PFS VDS and AutoWave can be linked together to a fully automotive test set up The set up communicates via the IEEE GPIB bus and is controlled by...

Страница 30: ...AMETEK CTS VDS 200Q10 Series Manual for Operation V 1 00 30 35 5 4 Example Test setup with VDS 200Q10 x AutoWave and PFM 200N100 Setup 12 example with Rack AutoWave PFM 200N100 VDS 200Q10 x...

Страница 31: ...AMETEK CTS VDS 200Q10 Series Manual for Operation V 1 00 31 35 5 5 Example Test setup with AutoWave and VDS 200Q10 x Setup 13 example with Rack AutoWave VDS 200Q10 x Exernal PFM 200N100...

Страница 32: ...t equipment In reply to all these questions we like to approach this issue as follows AMETEK CTS make use of a solid state semiconductor switch technique to generate high voltage transients A precious...

Страница 33: ...red only once if several devices are ordered The delivered packing list is in each case valid for the delivery 6 2 Accessories and options User software iso control Test analysis and documentation wit...

Страница 34: ...1 2011 Safety requirements for electrical equipment for measurement control and laboratory use EMC Directive 2014 30 EU Standard s to which conformity is declared EN 61326 1 2013 Electrical equipment...

Страница 35: ...ive 2014 35 EU Standard to which conformity is declared EN 61010 1 2011 Safety requirements for electrical equipment for measurement control and laboratory use EMC Directive 2014 30 EU Standard s to w...

Отзывы: