4.
Technical Data
4.1. Test level
Voltage for VDS 200Q10.x
U = -60 to +60 V ± 2% with 0.1 V steps setting
Output resolution Front panel setting
internal ADC 12 Bit: approx. 0.24 V at range -60
…+60 V
Output resolution AutoWave setting
AutoWave ADC 16 Bit: Resolution < 1 mV
Current
VDS 200Q10.x-100
I = 0 A
– ± 10 A ± 5%
max. 10 A
VDS 200Q10.x-120
I = 0 A
– ± 10 A ± 5%
max. 10 A
VDS 200Q10.x-230
I = 0 A
– ± 10 A ± 5%
max. 10 A
Bandwidth (-3dB)
Rise Time
Independent of programmed voltage and voltage range.
DC-180 kHz
Typ. <3us (High Freq.), <10 us (Standard)
Source impedance
Zi = < 10 m
Zi = Ri DC ... 400 Hz
Recovery time t
rec
t
rec
= 90 % of max. excursion within 25
μs
Current limiter I
limit
I
limit
= 0...Imax
Voltage ripple U
r
Pulses
U
r
= < 0.2 V peak to peak
as per test routines
4.2. Trigger
Automatic
Auto release with preselected parameters
Manual
Manual release of a single event
Extern
External release by external trigger
Repetition rate
10 ms
– 99 s
Drop out duration td
10
s to 9900 ms
Dip duration
10
s to 9900 ms
4.3. Input/output
Test / - output
Safety laboratory connectors at front panel
High current connectors at rear panel
AUX IN
Safety laboratory connectors at rear panel
Analog input
-10 to +10 V / 10k
/ DC
– 150 kHz
External trigger input
5
– 15 V TTL signal (BNC connector)
CRO trigger output
5 V TTL signal (BNC connector)
4.4. Interfaces
Serial interface USB
Baudrate: Setting; 1,200 to 19,200 Baud
Parallel interface
IEEE; addresses 1 - 30
FRAME BUS
EM Test internal communication interface for certain generators