AMETEK CTS
VDS 200Q10 Series
Manual for Operation
V 1.00
16 / 35
3.3.2.5. Sinus Sweep
This test simulates a residual a.c. on the dc supply
During Int f3 is applied.
Limitations
Vb + Vp <= 60.0V
Vb - Vp
= 0.0V
Parameters:
Vb
0.0V
-
60.0V
Vp
0.0V
-
0.1
– 60V
f1
0.001Hz
-
180.000kHz
f2
0.001Hz
-
180.000kHz
f3
0.001Hz
-
180.000kHz
t1
0.1s
-
999.9s
t2
0.1s
-
999.9s
int
0s
-
999s
n
1
-
30,000 / endl.
tri
Auto / Manual
I
1A
-
(Imax)
3.3.2.6. Overvoltage Vmax
This test simulates a high energy load dump pulse.
Input restrictions
0.0 V
V1 + V2
60.0V
Parameters:
V1
0.0V
-
60.0V
V2
0.0V
-
60.0V
t1
0.01s
-
999.99 s
t2
0.01s
-
999.99 s
t3
0.01s
-
999.99 s
int
0.1s
-
99.9s
n
1
-
30,000 / endl.
tri
Auto / Manual
I
1A
-
(Imax)