The reduction of iodine in the sensor generates a current directly proportional to the
diffusion rate of iodine through the membrane, which is directly proportional to the
concentration of iodine in the sample. Because the iodine concentration depends on the
amount of total chlorine in the sample, the sensor current is ultimately proportional to the
total chlorine concentration.
The permeability of the membrane to iodine is a function of temperature. A Pt100 RTD in
the sensor measures the temperature, and the transmitter uses the temperature to
compensate the total chlorine reading for changes in membrane permeability.
Sensor maintenance is fast and easy. Replacing the membrane requires no special tools or
fixtures. Simply place the membrane assembly on the cathode and screw the retainer in
place. Installing a new membrane and replenishing the electrolyte takes only a few
minutes.
1.2
Specifications
Sample Conditioning System
Table 1-1:
Physical characteristics
Specifications
General
Enclosure
Fiberglass reinforced polyester, NEMA 3 (IP53)
suitable for marine environments
Dimensions
14.5 x 13.0 x 8.6 in. (369 x 329 x 218 mm)
Mounting
Wall
Ambient temperature
0 to 50 °C (32 to122 °F)
Ambient humidity
0 to 90% (non-condensing)
Power
115 Vac, 6.9 W, 50/60 Hz
230 Vac, 7.0 W, 50/60 Hz
Hazardous location
The TCL Sample Conditioning System has no
hazardous location approvals.
Pumps
EN 809:1998
Weight/shipping weight
14 lb/16 lb (6.5 kg/7.5 kg)
Sample requirements
Inlet connection
Compression fitting, accepts 1/4 in. OD tubing
Drain connection
3/4 in. barbed fitting (must drain to open at-
mosphere)
Inlet pressure
<100 psig (791 kPa abs)
Flow
At least 0.25 gph (15 mL/min)
Temperature
0 to 50 °C (32 to122 °F)
Total alkalinity
<300 mg/L as CaCO
3
. For samples containing
<50 mg/L alkalinity, consult the factory.
Description and specifications
2
Rosemount TCL
Содержание Rosemount TCL
Страница 4: ......
Страница 8: ...Contents iv Rosemount TCL ...
Страница 16: ...Description and specifications 8 Rosemount TCL ...
Страница 18: ...Principles of operation 10 Rosemount TCL ...
Страница 23: ...TCL Case Dimensions Figure 3 4 Reagent Tubing Assembly Figure 3 5 Installation Instruction Manual 15 ...
Страница 26: ...Installation 18 Rosemount TCL ...
Страница 34: ...Wiring 26 Rosemount TCL ...
Страница 46: ...Display and operation 38 Rosemount TCL ...
Страница 66: ...Programming the transmitter 58 Rosemount TCL ...
Страница 68: ...Digital communications 60 Rosemount TCL ...
Страница 106: ...Troubleshooting 98 Rosemount TCL ...