a dual input transmitter, temperature compensation items appear for the other
sensor. Only four items are shown at a time. To view the remaining items, scroll to
the bottom of the screen and continue scrolling.
4.
To make a change, move the cursor to the desired line and press ENTER.
A screen appears in which the present setting can be edited.
5.
To store a setting, press ENTER.
For an explanation of terms, see
6.
To return to the main display, press MENU and then EXIT.
7.7
Configuring security settings
7.7.1
Purpose
This section describes how to set security codes. There are three levels of security.
1.
A user can view the default display and diagnostic screens only.
2.
A user has access to the calibration and hold menus only.
3.
A user has access to all menus.
The security code is a three digit number. The table shows what happens when different
security codes (
XXX
and
YYY
) are assigned to
Calibration/Hold
and
All
.
000
means no security.
Calibration/Hold
All
What happens
000
XXX
User enters
XXX
and has access to all
menus.
XXX
YYY
User enters
XXX
and has access to Cali-
bration and Hold menus only. User en-
ters
YYY
and has access to all menus.
XXX
000
User needs no security code to have
access to all menus.
000
000
User needs no security code to have
access to all menus.
Programming the transmitter
Instruction Manual
55
Содержание Rosemount TCL
Страница 4: ......
Страница 8: ...Contents iv Rosemount TCL ...
Страница 16: ...Description and specifications 8 Rosemount TCL ...
Страница 18: ...Principles of operation 10 Rosemount TCL ...
Страница 23: ...TCL Case Dimensions Figure 3 4 Reagent Tubing Assembly Figure 3 5 Installation Instruction Manual 15 ...
Страница 26: ...Installation 18 Rosemount TCL ...
Страница 34: ...Wiring 26 Rosemount TCL ...
Страница 46: ...Display and operation 38 Rosemount TCL ...
Страница 66: ...Programming the transmitter 58 Rosemount TCL ...
Страница 68: ...Digital communications 60 Rosemount TCL ...