9.3
Calibrating total chlorine
9.3.1
Purpose
The continuous determination of total chlorine requires two steps. See
sample flows into a conditioning system (the Rosemount TCL) where it is treated with
acetic acid (vinegar) and potassium iodide. The acid lowers the pH, which allows total
chlorine in the sample to quantitatively oxidize the iodide to iodine. The treated sample
then flows to the sensor. The sensor is a membrane-covered amperometric sensor, whose
output is proportional to the concentration of iodine. Because the concentration of iodine
is also proportional to the concentration of total chlorine, the transmitter can be calibrated
to read total chlorine.
Determination of Total Chlorine
Figure 9-1:
shows a typical calibration curve for a total chlorine sensor. Because the sensor
really measures iodine, calibrating the sensor requires exposing it to a solution containing
no iodine (zero standard) and to a solution containing a known amount of iodine (full-scale
standard).
Sensor Current as a Function of Total Chlorine Concentration
Figure 9-2:
Calibration
Instruction Manual
63
Содержание Rosemount TCL
Страница 4: ......
Страница 8: ...Contents iv Rosemount TCL ...
Страница 16: ...Description and specifications 8 Rosemount TCL ...
Страница 18: ...Principles of operation 10 Rosemount TCL ...
Страница 23: ...TCL Case Dimensions Figure 3 4 Reagent Tubing Assembly Figure 3 5 Installation Instruction Manual 15 ...
Страница 26: ...Installation 18 Rosemount TCL ...
Страница 34: ...Wiring 26 Rosemount TCL ...
Страница 46: ...Display and operation 38 Rosemount TCL ...
Страница 66: ...Programming the transmitter 58 Rosemount TCL ...
Страница 68: ...Digital communications 60 Rosemount TCL ...
Страница 106: ...Troubleshooting 98 Rosemount TCL ...