9.4
Calibrating analog outputs
9.4.1
Purpose
Although the transmitter analog outputs are calibrated at the factory, they can be
trimmed in the field to match the reading from a standard milliameter. Both the low (0 or
4 mA) and high (20 mA) outputs can be trimmed.
9.4.2
Procedure
Complete the following steps to calibrate the transmitter's analog outputs.
1.
Connect a calibrated milliameter across the output you wish to calibrate. If a load is
already connected to the output, disconnect the load.
Do not put the milliameter in parallel with the load.
2.
Press MENU.
The main Menu screen appears. The cursor is on
Calibrate
.
3.
Press ENTER.
4.
Choose the output you wish to calibrate.
The transmitter simulates the low output current.
5.
Change the value in the display to match the reading from the milliameter.
The transmitter simulates the 20 mA output current.
Calibration
Instruction Manual
69
Содержание Rosemount TCL
Страница 4: ......
Страница 8: ...Contents iv Rosemount TCL ...
Страница 16: ...Description and specifications 8 Rosemount TCL ...
Страница 18: ...Principles of operation 10 Rosemount TCL ...
Страница 23: ...TCL Case Dimensions Figure 3 4 Reagent Tubing Assembly Figure 3 5 Installation Instruction Manual 15 ...
Страница 26: ...Installation 18 Rosemount TCL ...
Страница 34: ...Wiring 26 Rosemount TCL ...
Страница 46: ...Display and operation 38 Rosemount TCL ...
Страница 66: ...Programming the transmitter 58 Rosemount TCL ...
Страница 68: ...Digital communications 60 Rosemount TCL ...
Страница 106: ...Troubleshooting 98 Rosemount TCL ...