WARNING
ELECTRIC SHOCK
Attaching cables and wiring harnesses longer than 9.8 ft. (3 m) may cause injury to
personnel.
The analyzer passed electromagnetic compatibility (EMC) tests based on all electrical
cables and harnesses attached to it being 9.8 ft. (3 m) long.
2.4
Test connections
Before sealing the cable glands, it is vital to test the new connections in order that any
faults can be remedied before the cables are permanently sealed.
WARNING
EXPLOSION HAZARD
In accordance with IEC 60079-0: 2011 Clause 30.
Failure to observe this warning could cause an explosion or potentially hazardous situation
which, if not avoided, may cause death or personal injury.
The cable glands used in this analyzer are only suitable for use in areas with a low risk of
mechanical damage. Protect the cable glands suitably.
Temporarily seal the cable glands with a suitable sealant to allow testing.
The tests described below are a basic check to prove function, rather than a test of sensor
capability.
To start the analyzer, follow the steps in
.
WARNING
INSTALLER RESPONSIBILITY
In accordance with IEC 60079-0: 2011 Clause 30
Failure to observe this warning could cause a potentially hazardous situation which, if not
avoided, may cause death or personal injury.
It is the installer's responsibility to connect the analyzer to a suitable alarm or shutdown
facility.
2.4.1
Testing power input cables and circuit breakers
The power input cables and circuit breakers can be tested as follows.
NOTICE
This section is in accordance with IEC 60079-0: 2011 Clause 30. This section must not be
changed, amended, or removed.
Reference Manual
Install
00809-0100-5800
September 2019
Rosemount CT5800
51
Содержание Rosemount CT5800
Страница 1: ...Reference Manual 00809 0100 5800 Rev AA September 2019 Rosemount CT5800 Continuous Gas Analyzer ...
Страница 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Страница 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Страница 46: ...Figure 2 6 Earth Bond Install Reference Manual September 2019 00809 0100 5800 46 Emerson com Rosemount ...
Страница 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Страница 70: ...Operating the analyzer Reference Manual September 2019 00809 0100 5800 70 Emerson com Rosemount ...
Страница 80: ...Verifying gas concentrations Reference Manual September 2019 00809 0100 5800 80 Emerson com Rosemount ...
Страница 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Страница 162: ...Troubleshooting and diagnostics Reference Manual September 2019 00809 0100 5800 162 Emerson com Rosemount ...
Страница 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Страница 170: ...Preventative maintenance Reference Manual September 2019 00809 0100 5800 170 Emerson com Rosemount ...
Страница 176: ...ATEX IECEx Assessment Report Summary Reference Manual September 2019 00809 0100 5800 176 Emerson com Rosemount ...
Страница 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Страница 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...
Страница 187: ...Engineering Drawings Reference Manual September 2019 00809 0100 5800 182 Emerson com Rosemount ...