DANGER
ELECTRIC SHOCK
This article is in accordance with IEC 60079-0: 2011 Clause 30.
Failure to observe this precaution may cause death, injury, and/or damage to persons
and/or property.
The analyzer operates using mains voltage that is dangerous to life.
Make sure that the circuit breakers are set to OFF and tagged off before opening the
front cover.
In accordance with IEC 60079-0:2011 Clause 30, install the analyzer in a suitable position
with shading to protect it from the elements. The displays and control panel on the front
of the housing must not be exposed to direct UV light sources or direct sunlight.
WARNING
POTENTIAL ELECTROSTATIC CHARGING HAZARD
This article is in accordance with IEC 60079-0: 2011 Clause 30.
The keypad is non-conducting and may generate an ignition capable of electrostatic
charges under certain extreme conditions.
Ensure that the equipment is not installed in a location where it may be subjected to
external conditions (such as high pressure steam), which might cause a build-up of
electrostatic charges on non-conducting surfaces.
Only clean the analyzer with a damp cloth.
The analyzer has a T4 temperature classification which specifies the maximum surface
temperature of the instrument, under a fault condition, is 275 °F (135 °C). You must
ensure that no combustible gas concentrations are present, whether on a continual or
occasional basis, which have an ignition temperature below the T classification of the unit.
WARNING
FIRE AND EXPLOSION
Failure to observe this precaution may cause death, personal injury, and/or damage to
equipment.
Do not open the Ex d enclosure of the analyzer unless the atmosphere of the area is known
to be below the ignitable concentration of combustible gases or materials.
In accordance with IEC 60079-0: 2011 Clause 30, install the analyzer in a suitable position
with shading to protect it from the elements. Do not expose the displays and control panel
on the front of the housing to direct UV light sources or direct sunlight.
Provide a fuse at the installation site which will completely disconnect the analyzer in case
of failure. Installing an isolating switch may also be beneficial. In either case, these
components must be constructed to conform to recognized norms.
In accordance with IEC 60664-1, install the analyzer in an area of not more than Pollution
Degree 2.
Install
Reference Manual
September 2019
00809-0100-5800
36
Emerson.com/Rosemount
Содержание Rosemount CT5800
Страница 1: ...Reference Manual 00809 0100 5800 Rev AA September 2019 Rosemount CT5800 Continuous Gas Analyzer ...
Страница 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Страница 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Страница 46: ...Figure 2 6 Earth Bond Install Reference Manual September 2019 00809 0100 5800 46 Emerson com Rosemount ...
Страница 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Страница 70: ...Operating the analyzer Reference Manual September 2019 00809 0100 5800 70 Emerson com Rosemount ...
Страница 80: ...Verifying gas concentrations Reference Manual September 2019 00809 0100 5800 80 Emerson com Rosemount ...
Страница 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Страница 162: ...Troubleshooting and diagnostics Reference Manual September 2019 00809 0100 5800 162 Emerson com Rosemount ...
Страница 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Страница 170: ...Preventative maintenance Reference Manual September 2019 00809 0100 5800 170 Emerson com Rosemount ...
Страница 176: ...ATEX IECEx Assessment Report Summary Reference Manual September 2019 00809 0100 5800 176 Emerson com Rosemount ...
Страница 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Страница 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...
Страница 187: ...Engineering Drawings Reference Manual September 2019 00809 0100 5800 182 Emerson com Rosemount ...