Figure 7-45: Analysis cell fillets
CAUTION
Take care not to put any pressure on or cause damage to the TEC board.
6. Use a flat head screwdriver to remove the four protective caps.
7. Remove the four socket head cap screws.
8. Carefully lift the cell assembly out of the enclosure.
9. Inspect the replacement analysis cell for damage. If the parts are damaged, contact
your local Customer Care representative.
10. Insert the replacement analysis cell into the enclosure.
11. Secure the four socket head screws and spring washers.
12. Attach the four protective caps.
13. Attach the DIN rail with two M5 socket head cap screws, M5 Spring washers, and
M5 plain washers securing the DIN rail support bracket to the base of the enclosure.
14. Connect all the component cables and connections on the lower DIN rail.
15. Connect the two 14 mm gas input and output line connectors at the flame arrestor.
Replacing the gas input and output lines
16. Close the enclosure as described in
and secure with captive
M16 bolts and engage the threads on the rear housing.
17. Remove the lock-out tag-out tag labels.
Reference Manual
Troubleshooting and diagnostics
00809-0100-5800
September 2019
Rosemount CT5800
159
Содержание Rosemount CT5800
Страница 1: ...Reference Manual 00809 0100 5800 Rev AA September 2019 Rosemount CT5800 Continuous Gas Analyzer ...
Страница 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Страница 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Страница 46: ...Figure 2 6 Earth Bond Install Reference Manual September 2019 00809 0100 5800 46 Emerson com Rosemount ...
Страница 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Страница 70: ...Operating the analyzer Reference Manual September 2019 00809 0100 5800 70 Emerson com Rosemount ...
Страница 80: ...Verifying gas concentrations Reference Manual September 2019 00809 0100 5800 80 Emerson com Rosemount ...
Страница 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Страница 162: ...Troubleshooting and diagnostics Reference Manual September 2019 00809 0100 5800 162 Emerson com Rosemount ...
Страница 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Страница 170: ...Preventative maintenance Reference Manual September 2019 00809 0100 5800 170 Emerson com Rosemount ...
Страница 176: ...ATEX IECEx Assessment Report Summary Reference Manual September 2019 00809 0100 5800 176 Emerson com Rosemount ...
Страница 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Страница 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...
Страница 187: ...Engineering Drawings Reference Manual September 2019 00809 0100 5800 182 Emerson com Rosemount ...