Table A-4: North America
(continued)
Detail
CSA North American Marking
Serial number
As applicable
Year of manufacture
As applicable
Warnings
WARNING - DO NOT OPEN WHEN AN EXPLOSIVE ATMOSPHERE IS
PRESENT.
WARNING - POTENTIAL ELECTROSTATIC CHARGING HAZARD - SEE
INSTRUCTIONS.
WARNING – DO NOT OPEN WHEN ENERGIZED.
WARNING – A SEAL SHALL BE INSTALLED WITHIN 50 mm OF THE
ENCLOSURE.
Maximum supply pressure
1 Bar
A.9
Conditions of certification/special conditions
for safe use
•
This equipment has flamepaths which differ from those in IEC60079-1/EN 60079-1.
Cascade Technologies Ltd. shall be contacted for guidance when maintaining the
flamepaths.
•
The fasteners which secure the cover are non-standard and shall therefore only be
replaced by fasteners supplied by the manufacturer for this purpose. The fasteners
must always be fitted with the washer supplied by the manufacturer.
•
The equipment has non-conductive surfaces which are a potential electrostatic
charging hazard - see instructions for guidance.
•
The user shall ensure that the flow process gas is limited to a maximum flow rate of 6
liters per minute.
•
The equipment shall only be uses with process gases which are classified for equipment
group IIB + H2 and must not contain oxygen or any other oxidizer in concentrations
greater than those found in normal air.
North American conditions
•
The equipment has flameproof joints with dimensions which are other than those
specified in Table 2 of ANSI/UL 60079-1: 6th edition and Table 3 of CSA C22.2
60079-1:16. These flameproof joints are not intended to be repaired but where
necessary the original manufacturer shall be contacted for guidance and information
on the dimensions of the flameproof joints.
•
The fasteners which secure the cover are non-standard and shall therefore only be
replaced by fasteners supplied by the manufacturer for this purpose. The fasteners
must always be fitted with the washer supplied by the manufacturer.
•
The equipment has non-conductive surfaces which are a potential electrostatic
charging hazard - see the instructions for guidance.
ATEX / IECEx Assessment Report Summary
Reference Manual
September 2019
00809-0100-5800
174
Emerson.com/Rosemount
Содержание Rosemount CT5800
Страница 1: ...Reference Manual 00809 0100 5800 Rev AA September 2019 Rosemount CT5800 Continuous Gas Analyzer ...
Страница 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Страница 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Страница 46: ...Figure 2 6 Earth Bond Install Reference Manual September 2019 00809 0100 5800 46 Emerson com Rosemount ...
Страница 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Страница 70: ...Operating the analyzer Reference Manual September 2019 00809 0100 5800 70 Emerson com Rosemount ...
Страница 80: ...Verifying gas concentrations Reference Manual September 2019 00809 0100 5800 80 Emerson com Rosemount ...
Страница 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Страница 162: ...Troubleshooting and diagnostics Reference Manual September 2019 00809 0100 5800 162 Emerson com Rosemount ...
Страница 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Страница 170: ...Preventative maintenance Reference Manual September 2019 00809 0100 5800 170 Emerson com Rosemount ...
Страница 176: ...ATEX IECEx Assessment Report Summary Reference Manual September 2019 00809 0100 5800 176 Emerson com Rosemount ...
Страница 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Страница 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...
Страница 187: ...Engineering Drawings Reference Manual September 2019 00809 0100 5800 182 Emerson com Rosemount ...