6
Gas calibration
6.1
Required tools
The gas concentrations measured by the analyzer can be validated against a known
sample gas or calibrated to match it by using the following gas calibration procedure.
To calibrate the analyzer, you need the following items:
•
Nitrogen gas of "instrument gas" purity for use as a zero calibration gas
•
Suitable span calibration gases for each gas measured
•
Gas bottle pressure regulators
•
Interconnecting hoses to connect the gas bottles to the analyzer
•
A T-piece and excess flow line
NOTICE
In the case of gases such as H
2
O, where it is not normally possible to obtain calibrated gas
cylinders, it will usually be measured by the same laser as some other gas. Validating the
other gases measured by the analyzer (particularly any which are measured by the same
laser as H
2
O) can demonstrate that the system is functioning correctly, meaning there is
no need to calibrate the water measurement directly.
If you need to calibrate (e.g., for legal requirements), you can use a water vapor generator
to supply a known concentration of water vapor.
6.2
Main menu
Access the calibration functions through the
Main menu
(
). To get to the
Main
menu
, press
MENU
on either the
Gas Sensor Main
screen or the
Pressure and
Temperature
Reference Manual
Gas calibration
00809-0100-5800
September 2019
Rosemount CT5800
81
Содержание Rosemount CT5800
Страница 1: ...Reference Manual 00809 0100 5800 Rev AA September 2019 Rosemount CT5800 Continuous Gas Analyzer ...
Страница 20: ...Contents Reference Manual September 2019 00809 0100 5800 xx Emerson com Rosemount ...
Страница 34: ...Plan Reference Manual September 2019 00809 0100 5800 34 Emerson com Rosemount ...
Страница 46: ...Figure 2 6 Earth Bond Install Reference Manual September 2019 00809 0100 5800 46 Emerson com Rosemount ...
Страница 60: ...Startup procedure Reference Manual September 2019 00809 0100 5800 60 Emerson com Rosemount ...
Страница 70: ...Operating the analyzer Reference Manual September 2019 00809 0100 5800 70 Emerson com Rosemount ...
Страница 80: ...Verifying gas concentrations Reference Manual September 2019 00809 0100 5800 80 Emerson com Rosemount ...
Страница 90: ...Gas calibration Reference Manual September 2019 00809 0100 5800 90 Emerson com Rosemount ...
Страница 162: ...Troubleshooting and diagnostics Reference Manual September 2019 00809 0100 5800 162 Emerson com Rosemount ...
Страница 168: ...Shutdown procedure Reference Manual September 2019 00809 0100 5800 168 Emerson com Rosemount ...
Страница 170: ...Preventative maintenance Reference Manual September 2019 00809 0100 5800 170 Emerson com Rosemount ...
Страница 176: ...ATEX IECEx Assessment Report Summary Reference Manual September 2019 00809 0100 5800 176 Emerson com Rosemount ...
Страница 178: ...Theory of operation Reference Manual September 2019 00809 0100 5800 178 Emerson com Rosemount ...
Страница 180: ...Spare parts list Reference Manual September 2019 00809 0100 5800 180 Emerson com Rosemount ...
Страница 187: ...Engineering Drawings Reference Manual September 2019 00809 0100 5800 182 Emerson com Rosemount ...