5
Program the transmitter
5.1
General programming information
Typical programming steps include the following listed procedures. You can easily and
quickly accomplish each of these programming functions using the intuitive menu system.
•
Configure and assign values to the current outputs.
•
Set a security code for two levels of security access.
•
Access menu functions using a security code.
•
Enable and disable Hold mode for current outputs.
•
Choose the frequency of the AC power (needed for optimum noise rejection).
•
Reset all factory defaults, calibration data only, or current output settings only.
5.2
Configuring and ranging current outputs
The transmitter accepts inputs from two sensors and has four analog current outputs.
Ranging the outputs means assigning values to the low (0 or 4 mA) and high (20 mA)
outputs.
Important
Always configure the outputs first.
To configure the outputs, press
ENTER/MENU
to go to
Program
→
Outputs
. You can set
Outputs 1 - 4 to Analog, PID, or Simulation. You can also set output assignments, scaling
the range, linear or logarithmic outputs, output dampening, output setpoints, and output
fault levels for each output if desired.
Reference Manual
Program the transmitter
00809-0100-3565
July 2020
Rosemount T56
35
Содержание Rosemount Clarity II T56
Страница 1: ...Reference Manual 00809 0100 3565 Rev AB July 2020 Rosemount Clarity II T56 Turbidmeter ...
Страница 4: ...4 ...
Страница 12: ...Description and specifications Reference Manual July 2020 00809 0100 3565 12 Emerson com Rosemount ...
Страница 16: ...Figure 2 4 Panel Cut Out A Maximum Install Reference Manual July 2020 00809 0100 3565 16 Emerson com Rosemount ...
Страница 20: ...Figure 2 8 Pipe Mount Side View Install Reference Manual July 2020 00809 0100 3565 20 Emerson com Rosemount ...
Страница 30: ...Wire Reference Manual July 2020 00809 0100 3565 30 Emerson com Rosemount ...
Страница 34: ...Display and operate Reference Manual July 2020 00809 0100 3565 34 Emerson com Rosemount ...
Страница 44: ...Program turbidity Reference Manual July 2020 00809 0100 3565 44 Emerson com Rosemount ...
Страница 48: ...Calibrate Reference Manual July 2020 00809 0100 3565 48 Emerson com Rosemount ...
Страница 56: ...Maintenance Reference Manual July 2020 00809 0100 3565 56 Emerson com Rosemount ...