7
Calibrate
7.1
Introduction to calibration
Calibration is the process of adjusting or standardizing the transmitter to a lab test or a
calibrated laboratory instrument or standardizing to some known reference.
Completing Quick Start upon first power-up enables live measurements but does not
ensure accurate readings in the lab or in process. Calibrate each attached sensor to ensure
accurate, repeatable readings.
7.2
Turbidity
This section describes how to calibrate the turbidity sensor against a user-prepared
standard as a two-point calibration with deionized water, against a 20 NTU user-prepared
standard as a single point calibration, and against a grab sample using a reference
turbidmeter.
To calibrate the turbidity sensor:
1. Press
ENTER/MENU
from the main screen to access the
Calibration
screen.
Figure 7-1: Calibration Screen
2. Select S1 Measurement or S2 Measurement and press
ENTER/MENU
.
3. Press
INFO
at any time to learn more about this procedure. When you press
INFO
, a
yellow screen appears with detailed instructions and information.
The following calibration routines are covered:
•
Slope Calibration: Slope calibration with pure water and a standard of known turbidity.
Reference Manual
Calibrate
00809-0100-3565
July 2020
Rosemount T56
45
Содержание Rosemount Clarity II T56
Страница 1: ...Reference Manual 00809 0100 3565 Rev AB July 2020 Rosemount Clarity II T56 Turbidmeter ...
Страница 4: ...4 ...
Страница 12: ...Description and specifications Reference Manual July 2020 00809 0100 3565 12 Emerson com Rosemount ...
Страница 16: ...Figure 2 4 Panel Cut Out A Maximum Install Reference Manual July 2020 00809 0100 3565 16 Emerson com Rosemount ...
Страница 20: ...Figure 2 8 Pipe Mount Side View Install Reference Manual July 2020 00809 0100 3565 20 Emerson com Rosemount ...
Страница 30: ...Wire Reference Manual July 2020 00809 0100 3565 30 Emerson com Rosemount ...
Страница 34: ...Display and operate Reference Manual July 2020 00809 0100 3565 34 Emerson com Rosemount ...
Страница 44: ...Program turbidity Reference Manual July 2020 00809 0100 3565 44 Emerson com Rosemount ...
Страница 48: ...Calibrate Reference Manual July 2020 00809 0100 3565 48 Emerson com Rosemount ...
Страница 56: ...Maintenance Reference Manual July 2020 00809 0100 3565 56 Emerson com Rosemount ...