Inputs
One or two isolated sensor inputs.
Outputs
Four 4-20 mA or 0-20 mA isolated current outputs. Fully
scalable. Maximum load: 550 ohm. Output 1 has superimposed
HART
®
signal.
Output dampening
0-999 seconds
Current output
accuracy
±0.05 mA at 77 °F (25 °C)
Alarms
Four process alarm relays for turbidity or temperature. You can
also program relays for timer, TPC, or fault alarm operation
instead of as process alarms. You can configure each relay
independently. Alarm logic (high or low activation) and
deadband are user-programmable.
Terminal connections
rating
Power connector (3 leads): 24-12 AWG wire size.
Signal board terminal blocks: 26-16 AWG wire size.
Current output connectors (2 leads): 24-16 AWG wire size.
Alarm relay terminal blocks: 24-12 AWG wire size.
RFI/EMI
EN 61326
LVD
EN-61010-1
Relays
Form C, single pole, double throw, epoxy sealed
Table 1-1: Maximum Relay Current
Supply voltage
Resistive
28 Vdc 5.0 A
5.0 A
115 Vac 5.0 A
5.0 A
230 Vac 5.0 A
5.0 A
Field wiring terminals
Removable terminal blocks for power, analog outputs, and
sensors.
Transmitter hazardous location approvals
CSA
approvals
Class 1, Division 2, Groups A, B, C, and D
Class II, Division 2, Groups E, F, and G
Class III T4A Tamb = 50 °C
Evaluated to the ANS/UL standards. The C and US indicators adjacent to
the CSA mark signify that the product has been evaluated to the
applicable CSA and ANSI/UL standards for use in Canada and the US
respectively.
Reference Manual
Description and specifications
00809-0100-3565
July 2020
Rosemount T56
9
Содержание Rosemount Clarity II T56
Страница 1: ...Reference Manual 00809 0100 3565 Rev AB July 2020 Rosemount Clarity II T56 Turbidmeter ...
Страница 4: ...4 ...
Страница 12: ...Description and specifications Reference Manual July 2020 00809 0100 3565 12 Emerson com Rosemount ...
Страница 16: ...Figure 2 4 Panel Cut Out A Maximum Install Reference Manual July 2020 00809 0100 3565 16 Emerson com Rosemount ...
Страница 20: ...Figure 2 8 Pipe Mount Side View Install Reference Manual July 2020 00809 0100 3565 20 Emerson com Rosemount ...
Страница 30: ...Wire Reference Manual July 2020 00809 0100 3565 30 Emerson com Rosemount ...
Страница 34: ...Display and operate Reference Manual July 2020 00809 0100 3565 34 Emerson com Rosemount ...
Страница 44: ...Program turbidity Reference Manual July 2020 00809 0100 3565 44 Emerson com Rosemount ...
Страница 48: ...Calibrate Reference Manual July 2020 00809 0100 3565 48 Emerson com Rosemount ...
Страница 56: ...Maintenance Reference Manual July 2020 00809 0100 3565 56 Emerson com Rosemount ...