Install
2.1
Unpack and inspect
Consult the table to verify that you have received the parts for the option you ordered.
Table 2-1: Rosemount T56 Turbidmeter Parts
Item
Model/part number
Single input turbidity transmitter with HART
®
56-03-27-38-HT
Dual input turbidity transmitter with HART
56-03-27-37-HT
Sensor - EPA standards
8-0108-0002-EPA
Sensor - ISO standard
8-0108-0003-ISO
Cable - 3 ft. (0.9 m)
2413800
Cable - 20 ft. (6.1 m)
2409700
Cable - 50 ft. (15.2 m)
2409800
Calibration cup
2410100
Molded chamber/debubbler
24170-00
2.2
General installation information
1. Although the transmitter is suitable for outdoor use, do not install it in direct
sunlight or in areas of extreme temperatures.
2. Install the transmitter in an area where vibration and electromagnetic and radio
frequency interference are minimized or absent.
3. Keep the transmitter and sensor wiring at least one foot (0.3 m) from high voltage
conductors. Be sure there is easy access to the transmitter.
4. The transmitter is suitable for panel, pipe, or surface mounting. Refer to the figures
below.
WARNING
Electrical shock
Electrical installation must be in accordance with the National Electrical Code (ANSI/
NFPA-70) and/or any other applicable national or local codes.
Do not operate or energize instrument with case open.
Note
Dimensions are in inches (millimeters).
1. Panel supplied by others. Maximum thickness: 3.75 in. (9.52 mm)
2. 4X mounting brackets and screws provided with instrument
Reference Manual
Install
00809-0100-3565
July 2020
Rosemount T56
13
Содержание Rosemount Clarity II T56
Страница 1: ...Reference Manual 00809 0100 3565 Rev AB July 2020 Rosemount Clarity II T56 Turbidmeter ...
Страница 4: ...4 ...
Страница 12: ...Description and specifications Reference Manual July 2020 00809 0100 3565 12 Emerson com Rosemount ...
Страница 16: ...Figure 2 4 Panel Cut Out A Maximum Install Reference Manual July 2020 00809 0100 3565 16 Emerson com Rosemount ...
Страница 20: ...Figure 2 8 Pipe Mount Side View Install Reference Manual July 2020 00809 0100 3565 20 Emerson com Rosemount ...
Страница 30: ...Wire Reference Manual July 2020 00809 0100 3565 30 Emerson com Rosemount ...
Страница 34: ...Display and operate Reference Manual July 2020 00809 0100 3565 34 Emerson com Rosemount ...
Страница 44: ...Program turbidity Reference Manual July 2020 00809 0100 3565 44 Emerson com Rosemount ...
Страница 48: ...Calibrate Reference Manual July 2020 00809 0100 3565 48 Emerson com Rosemount ...
Страница 56: ...Maintenance Reference Manual July 2020 00809 0100 3565 56 Emerson com Rosemount ...