Verify linearity
This procedure describes how to verify linearity between turbidity and TSS.
Procedure
1. Collect a sample of the process liquid.
You may need 10 L or more if you use the Rosemount
™
Clarity II for measuring
turbidity. If you uses a laboratory turbidmeter, you will need less volume. The
Rosemount Clarity II requires about 500 mL for the measurement; laboratory
turbidmeters require 50 mL or less. Verify that the turbidity of the sample is less
than 200 NTU. Store in a clean bottle.
2. Filter a portion of the sample to obtain at least 5 L of dilution liquid.
The filtrate is needed to dilute the sample in subsequent steps. Verify that the
turbidity of the dilution water is low. If filtering the sample is impractical, use
deionized water for dilution.
3. Measure the total suspended solids (TSS) in the sample obtained in
Thoroughly mix the sample before withdrawing liquid.
A magnetic stirrer is best, but inverting the sample repeatedly for about a minute
works too. Avoid violent shaking or mixing. Refer to any standard reference work on
water and wastewater testing for the procedure for determining TSS.
by a factor of 0.9, 0.7, 0.5, 0.3, and 0.1.
See the table for recommended volumes. Measure TSS and turbidity for each
dilution. For lower TSS values, use a larger volume of sample.
Dilution factor
Volume of stock, mL
Final volume, mL
Volume for
Rosemount Clarity II,
mL
Volume for TSS, mL
1.00
--
--
500
50 - 250
0.9
900
1000
500
50 - 250
0.7
700
1000
500
50 - 250
0.5
500
1000
500
50 - 250
0.3
300
1000
500
50 - 250
0.1
100
1000
500
50 - 250
, with turbidity on the Y-axis and TSS on the X-axis.
Fit the best straight line to the data.
6. Locate two points (P1 and P2) on the line separated as much as possible. Read the
ppm and NTU value for each point and enter these into the transmitter.
Reference Manual
Verify linearity
00809-0100-3565
July 2020
Rosemount T56
59
Содержание Rosemount Clarity II T56
Страница 1: ...Reference Manual 00809 0100 3565 Rev AB July 2020 Rosemount Clarity II T56 Turbidmeter ...
Страница 4: ...4 ...
Страница 12: ...Description and specifications Reference Manual July 2020 00809 0100 3565 12 Emerson com Rosemount ...
Страница 16: ...Figure 2 4 Panel Cut Out A Maximum Install Reference Manual July 2020 00809 0100 3565 16 Emerson com Rosemount ...
Страница 20: ...Figure 2 8 Pipe Mount Side View Install Reference Manual July 2020 00809 0100 3565 20 Emerson com Rosemount ...
Страница 30: ...Wire Reference Manual July 2020 00809 0100 3565 30 Emerson com Rosemount ...
Страница 34: ...Display and operate Reference Manual July 2020 00809 0100 3565 34 Emerson com Rosemount ...
Страница 44: ...Program turbidity Reference Manual July 2020 00809 0100 3565 44 Emerson com Rosemount ...
Страница 48: ...Calibrate Reference Manual July 2020 00809 0100 3565 48 Emerson com Rosemount ...
Страница 56: ...Maintenance Reference Manual July 2020 00809 0100 3565 56 Emerson com Rosemount ...