6. Insert the replacement board in the sensor and push the socket on the replacement
board into the mating pins in the sensor.
CAUTION
Equipment damage
Keep wires pushed away from lamp board when replacing the flange.
A. Wires and flange
B. Mating pins
7. Place the dessicant package in the sensor body.
8. Orient the flange so that the screw holes line up with the holes in the sensor body.
Push the flange back on the sensor body and replace the screws.
Don't let wires push on lamp board. You may need to turn the flange a small
amount until the holes line up.
9. Place the sensor in the calibration cup and reconnect the cable.
10. Calibrate the sensor using either slope or standard calibration.
See
or
. Do not use grab calibration.
Failure to calibrate the sensor may reduce the life of the sensor.
Maintenance
Reference Manual
July 2020
00809-0100-3565
52
Emerson.com/Rosemount
Содержание Rosemount Clarity II T56
Страница 1: ...Reference Manual 00809 0100 3565 Rev AB July 2020 Rosemount Clarity II T56 Turbidmeter ...
Страница 4: ...4 ...
Страница 12: ...Description and specifications Reference Manual July 2020 00809 0100 3565 12 Emerson com Rosemount ...
Страница 16: ...Figure 2 4 Panel Cut Out A Maximum Install Reference Manual July 2020 00809 0100 3565 16 Emerson com Rosemount ...
Страница 20: ...Figure 2 8 Pipe Mount Side View Install Reference Manual July 2020 00809 0100 3565 20 Emerson com Rosemount ...
Страница 30: ...Wire Reference Manual July 2020 00809 0100 3565 30 Emerson com Rosemount ...
Страница 34: ...Display and operate Reference Manual July 2020 00809 0100 3565 34 Emerson com Rosemount ...
Страница 44: ...Program turbidity Reference Manual July 2020 00809 0100 3565 44 Emerson com Rosemount ...
Страница 48: ...Calibrate Reference Manual July 2020 00809 0100 3565 48 Emerson com Rosemount ...
Страница 56: ...Maintenance Reference Manual July 2020 00809 0100 3565 56 Emerson com Rosemount ...