2. Enter the three-digit security code for the appropriate security level.
If the entry is correct, the appropriate menu screen appears. If the entry is incorrect,
the
Invalid Code
screen appears. The
Enter security code
screen reappears after two
seconds.
5.4
Use hold
To prevent improper operation of systems or pumps that are controlled directly by the
current output, place the transmitter in hold before removing the sensor for calibration
and maintenance.
During hold, all outputs remain at the last value.
Important
Once in hold, all current outputs remain on hold indefinitely.
Reference Manual
Program the transmitter
00809-0100-3565
July 2020
Rosemount T56
37
Содержание Rosemount Clarity II T56
Страница 1: ...Reference Manual 00809 0100 3565 Rev AB July 2020 Rosemount Clarity II T56 Turbidmeter ...
Страница 4: ...4 ...
Страница 12: ...Description and specifications Reference Manual July 2020 00809 0100 3565 12 Emerson com Rosemount ...
Страница 16: ...Figure 2 4 Panel Cut Out A Maximum Install Reference Manual July 2020 00809 0100 3565 16 Emerson com Rosemount ...
Страница 20: ...Figure 2 8 Pipe Mount Side View Install Reference Manual July 2020 00809 0100 3565 20 Emerson com Rosemount ...
Страница 30: ...Wire Reference Manual July 2020 00809 0100 3565 30 Emerson com Rosemount ...
Страница 34: ...Display and operate Reference Manual July 2020 00809 0100 3565 34 Emerson com Rosemount ...
Страница 44: ...Program turbidity Reference Manual July 2020 00809 0100 3565 44 Emerson com Rosemount ...
Страница 48: ...Calibrate Reference Manual July 2020 00809 0100 3565 48 Emerson com Rosemount ...
Страница 56: ...Maintenance Reference Manual July 2020 00809 0100 3565 56 Emerson com Rosemount ...