113
Reference Manual
00809-0100-4570, Rev DA
Operation
November 2017
Operation
Figure 7-74. False Echo Example in Echo Curve Analysis Window
5. Zoom into the area of false echoes at the first 1.5 m by clicking and dragging down and to the right
marking a square around the required area. This is done to view the exact numbers that later needed
to be mapped out under the False Echo Mapping function.
6. Check the
Map Out False Echoes
check box, select the required tab or beam on which you would like to
perform the false echo mapping, using the mouse left click mark a rectangle around the false echo
you wish to map and release the mouse button. The
Echo Mapping
window will pop up populating all
relevant fields, Click
Start Scanning
to initiate the operation.
7. False Echo Mapping can be also found under the Device menu. Manually scan from 0 to beyond the
distance of the false echoes. In this case, the echo will be 1.6 m. Select Manual Scan, set the range of
scan and the level of the threshold, in this case it can be 10000.
8. Select the
Start Scanning
button.
9. Once the message disappears, the False Echo Mapping process is completed. Wait a couple of
minutes and perform Echo Curve Analysis again to verify the position of the User False Echo line (pink)
in each of the beams the operation was done. Wait at least one damping time before the algorithm
takes effect of the changes and completely ignores the false echoes.
Note
This is a brief run through of the False Echo Mapping process. There are many other situations that may
call for different techniques. An improper mapping of a vessel can lead to a poor performance.
Содержание Rosemount 5708
Страница 1: ...Reference Manual 00809 0100 4570 Rev DA November 2017 Rosemount 5708 3D Solids Scanner ...
Страница 2: ...ii Section Title November 2017 Reference Manual 00809 0100 4570 Rev DA ...
Страница 8: ...vi Reference Manual 00809 0100 4570 Rev DA Contents November 2017 Contents ...
Страница 12: ...2 Reference Manual 00809 0100 4570 Rev DA Introduction November 2017 Introduction ...
Страница 16: ...6 Reference Manual 00809 0100 4570 Rev DA Rosemount 5708 Overview November 2017 Rosemount 5708 Overview ...
Страница 42: ...32 Reference Manual 00809 0100 4570 Rev DA Electrical Installation November 2017 Electrical Installation ...
Страница 174: ...Connection via RS 485 November 2017 Reference Manual 00809 0100 4570 Rev DA Connection via RS 485 164 ...
Страница 202: ...ii Reference Manual 00809 0100 4570 Rev DA Index November 2017 Index ...
Страница 203: ...iii Reference Manual 00809 0100 4570 Rev DA November 2017 ...