DL8000 Preset Instruction Manual
Revised February-2016
DL8000 Keypad Display
E-53
14/03/12 12:26:02
For traversing use
UP | DOWN
LEFT | RIGHT
ENTER to continue
EXIT to Go Back
EXIT to quit
Additive Calib
Analog Calib
RTD Calib
PRODUCT Calib
(Pg:1/1)
E.4.2 Calibrating Analog Input
Note
:
The preset must have an installed AI module to calibrate AI. If
you attempt an AI calibration without a module, the system
displays the message
NO AI Card in system
on the left-hand
screen.
To calibrate analog inputs, select the
Analog Calib
option on the
Calibration menu screen:
14/03/12 12:26:02
For traversing use
UP | DOWN
LEFT | RIGHT
ENTER to continue
EXIT to Go Back
EXIT to quit
Additive Calib
Analog Calib
RTD Calib
PRODUCT Calib
(Pg:1/1)
The system then displays all available analog inputs. Select an analog
input channel for calibration:
14/03/12 12:26:02
Sel Analog Ch
Sel No of Mid pt
(SELECT)
AI ch 1
Then choose the number of mid-points. Press
SELECT
to see the
available options (zero, 1, 2, and 3):
14/03/12 12:26:02
Sel Analog Ch
Sel No of Mid pt
(SELECT)
AI ch 1
Zero