9. If required, install the wire seals through the socket head bolts on the Base
Enclosure (see
Sealing the Transmitter Electronics Enclosure
).
10. Connect electrical power to the system.
11. Set or configure the meter using Daniel MeterLink. For additional installation
information refer to the system wiring diagram (see
),
Daniel MeterLink
Softeware for Gas and Liquid Ultrasonic Meters Quick Start Manual
(P/N 3-9000-763).
Use the Daniel MeterLink Field Setup Wizard to complete the configuration. Also
see,
Field Setup Wizard using Daniel MeterLink
in this manual.
3.4.2
Startup for systems using flame-proof cable
WARNING
HAZARDOUS VOLTAGE INSIDE
Do not open the Transmitter Electronics Enclosure when an explosive gas atmosphere is
present. Disconnect equipment from supply circuit before opening.
Failure to remove power may result in death or serious injury.
Procedure
1. Check to make certain that all field wiring power is turned
OFF
.
2. Remove the end cap nearest the cable entries to gain access to the transmitter
electronics.
3. Install the cable and cable gland.
4. Complete the field connection wiring.
5. Connect a flow computer to the communications line on the Daniel 3814 Liquid
Ultrasonic Flow Meter.
6. Apply electrical power to the system to ensure the field connections are working
correctly. Allow the system to run for the time specified by the customer (usually
one week) and an electrician have fully tested the connections. After the
Acceptance Test is witnessed and approved, seal the conduit.
7. Connect electrical power to the system.
8. If required, install the security latches and wire seals on the Transmitter Electronics
Enclosure end caps .
9. Install the wire seals on the Base Enclosure hex head bolts and on the Shroud
covering the meter body (see
10. Ensure the safety latches are installed on the Transmitter Electronics Enclosure.
11. Set or configure the software using Daniel MeterLink. For additional installation
information refer to the system wiring diagram (see
),
Daniel MeterLink
Software for Gas and Liquid Quick Start Manual
(P/N 3-9000-763). Use the Daniel
MeterLink Field Setup Wizard to complete the meter configuration. Also see,
Setup Wizard using Daniel MeterLink
Electrical Installation
Installation manual
May 2019
3-9000-760
52
3814 Liquid Ultrasonic Flow Meter
Содержание Daniel 3814
Страница 1: ...Installation manual 3 9000 760 Rev HA May 2019 Daniel 3814 Liquid Ultrasonic Flow Meter ...
Страница 2: ...2 ...
Страница 10: ...Contents Installation manual May 2019 3 9000 760 x 3814 Liquid Ultrasonic Flow Meter ...
Страница 48: ...Mechanical Installation Installation manual May 2019 3 9000 760 48 3814 Liquid Ultrasonic Flow Meter ...
Страница 92: ...Configuration Installation manual May 2019 3 9000 760 92 3814 Liquid Ultrasonic Flow Meter ...
Страница 94: ......
Страница 95: ......
Страница 96: ......
Страница 97: ......
Страница 98: ...Engineering Drawings Installation manual May 2019 3 9000 760 94 3814 Liquid Ultrasonic Flow Meter ...