Appendix C
Surge Protector / C-1
Appendix C
SURGE PROTECTOR
Pt. No. 388630-01-9
Transients caused by power and lightning surges can cause damage to field-mounted units.
Transmitter models without meters have bipolar, differential surge protection rated at 1000
watts for 1 millisecond. Consistent surges that exceed these levels could cause a failure of
the Transmitter's surge protection diode.
Additional protection can be achieved through the use of external devices. The optional
Joslyn 1669-02 Transient Surge Protector (Bristol Babcock Pt. No. 388630-01-9) provides
protection for surges up to 10000 amperes. The protector diverts excess surge currents
around the Transmitter and maintains the internal current at a low level.
The Transient Surge Protector is wired as shown in the illustration. The end having the
three wire leads (red, black and green) will screw into one of the 1/2 inch NPT conduit
openings of the Ttransmitter. The red lead connects to , black to SIGNAL-, and
green to the green grounding screw. The earth ground lead should also be connected to the
grounding screw. The red and black leads at the other end of the surge protector connect to
the + and
−
side of the loop as shown.
The internal 44 ohms resistance of the Surge Protector must also be added to the loop
resistance when calculating the Transmitter's supply voltage.
Warning
The Surge Protector is not approved for operation in hazardous areas. Its use
is strictly limited to Transmitters that operate in non-hazardous installations.
Wiring Surge Protector Wiring Connections to 2808 Transmitter
Содержание Bristol 2808 Series
Страница 12: ...BLANK PAGE ...
Страница 14: ...1 2 Introduction 2808B Figure 1 1 2808 XXB Transmitters ...
Страница 16: ...BLANK PAGE ...
Страница 23: ...2808B GP Transmitters 1A 7 Figure 1A 5 Overall Dimensions Model 2808 15B With Neck Type Mounting Bracket ...
Страница 24: ...1A 8 GP Transmitters 2808B Figure 1A 6 Overall Dimensions Model 2808 16B With Neck Type Mounting Bracket ...
Страница 30: ...BLANK PAGE ...
Страница 32: ...1B 2 DP Transmitters 2808B Figure 1B 1 Model 2808 35B Figure 1B 2 Transmitter Assemblies ...
Страница 37: ...2808B DP Transmitters 1B 7 Figure 1B 4C Dimensions for Model 2808 35B Transmitter With Manifold Adapter Option ...
Страница 40: ...1B 10 DP Transmitters 2808B Figure 1B 6 Transmitter Housing Rotation Diagram ...
Страница 54: ...BLANK PAGE ...
Страница 57: ...2808B Calibration 3 3 Figure 3 2 Calibration Adjustments ...
Страница 60: ...3 6 Calibration 2808B TABLE 3 A COARSE ZERO SWITCH SETTINGS Elevation ...
Страница 61: ...2808B Calibration 3 7 TABLE 3 B COARSE ZERO SWITCH SETTINGS Suppression ...
Страница 66: ...BLANK PAGE ...
Страница 68: ...4 2 Service 2808B Figure 4 1 Using Internal TEST Terminals to Measure Current ...
Страница 72: ...BLANK PAGE ...
Страница 74: ...BLANK PAGE ...
Страница 81: ...Appendix B Page 5 Loop Powered Indicator Figure 2 2 Loop Powered Indicator Installation Diagram ...
Страница 90: ...BLANK PAGE ...
Страница 92: ...BLANK PAGE ...
Страница 100: ...BLANK PAGE ...
Страница 127: ...ESDS Manual S14006 4 15 92 CARE AND HANDLING OF PC BOARDS AND ESD SENSITIVE COMPONENTS BRISTOL BABCOCK ...
Страница 128: ...BLANK PAGE ...
Страница 139: ...BLANK PAGE ...