3-10 / Calibration
2808B
2.
Set the Dip and Rotary switches in accordance with Table 3-B. Suppressed zero values
are expressed as negative percentages of upper range limit (URL). Zero percent URL
represents no suppression. If the desired percent suppression is in between values, try
the next closest setting.
3.
Apply a 100% pressure input to the transmitter equivalent to the upper range value
(URV). For this example, URV = 80 psi.
4.
If Transmitter output is not 5.000V (20.00mA), adjust A1 to make a minor correction.
5.
Lower the input pressure to P
min
. Adjust A2 to set the output to 1.000V (4.00mA).
6.
Recheck the Transmitter output with 0%, and 100% inputs. The DMM should provide
respective readings of 1.000V (4.00mA), and 5.000V (20.00mA) (
±
0.15% full scale).
7.
If necessary, repeat the A1 and A2 adjustment procedures.
8.
If problems persist, recheck the DIP switch settings. Try setting the switches to the
lower percentage. For instance, if the desired percent suppression is -85%, set the
switches for a suppression of -90%.
Figure 3-3 - Water Tower Level Measurement - Zero Suppression
Содержание Bristol 2808 Series
Страница 12: ...BLANK PAGE ...
Страница 14: ...1 2 Introduction 2808B Figure 1 1 2808 XXB Transmitters ...
Страница 16: ...BLANK PAGE ...
Страница 23: ...2808B GP Transmitters 1A 7 Figure 1A 5 Overall Dimensions Model 2808 15B With Neck Type Mounting Bracket ...
Страница 24: ...1A 8 GP Transmitters 2808B Figure 1A 6 Overall Dimensions Model 2808 16B With Neck Type Mounting Bracket ...
Страница 30: ...BLANK PAGE ...
Страница 32: ...1B 2 DP Transmitters 2808B Figure 1B 1 Model 2808 35B Figure 1B 2 Transmitter Assemblies ...
Страница 37: ...2808B DP Transmitters 1B 7 Figure 1B 4C Dimensions for Model 2808 35B Transmitter With Manifold Adapter Option ...
Страница 40: ...1B 10 DP Transmitters 2808B Figure 1B 6 Transmitter Housing Rotation Diagram ...
Страница 54: ...BLANK PAGE ...
Страница 57: ...2808B Calibration 3 3 Figure 3 2 Calibration Adjustments ...
Страница 60: ...3 6 Calibration 2808B TABLE 3 A COARSE ZERO SWITCH SETTINGS Elevation ...
Страница 61: ...2808B Calibration 3 7 TABLE 3 B COARSE ZERO SWITCH SETTINGS Suppression ...
Страница 66: ...BLANK PAGE ...
Страница 68: ...4 2 Service 2808B Figure 4 1 Using Internal TEST Terminals to Measure Current ...
Страница 72: ...BLANK PAGE ...
Страница 74: ...BLANK PAGE ...
Страница 81: ...Appendix B Page 5 Loop Powered Indicator Figure 2 2 Loop Powered Indicator Installation Diagram ...
Страница 90: ...BLANK PAGE ...
Страница 92: ...BLANK PAGE ...
Страница 100: ...BLANK PAGE ...
Страница 127: ...ESDS Manual S14006 4 15 92 CARE AND HANDLING OF PC BOARDS AND ESD SENSITIVE COMPONENTS BRISTOL BABCOCK ...
Страница 128: ...BLANK PAGE ...
Страница 139: ...BLANK PAGE ...