Appendix B
Page 3
Loop Powered Indicator
Section 2
INSTALLATION, OPERATION & SERVICE
2.1 INSTALLATION & REMOVAL/REPLACEMENT OF THE LPI
2.1.1 Installation/Removal of the Loop Powered Indicator
The following parts are provided for field installation of the Loop Powered Indicator option:
One (1) - O-Ring, Size - 149
Two (2) - Standoffs, #4 M/F Shoulder
Two (2) - Standoffs, 4-40 x .750 M/F
One (1) - Meter Motherboard Ass’y.
One (1) - Meter/Display Board Ass’y.
Two (2) - Standoffs, 4-40 x .375 M/F
Four (4) - Screws, 4-40 x 1⁄4 Pan Head
One (1) - 2808 Display Cover Ass’y.
WARNING
Never attempt to service a Model 2808 Transmitter while
it is operating in a hazardous environment. Either the
area must be made safe or the unit must be unwired,
unmounted, and taken to a safe, non-hazardous area.
WARNING
Never attempt to install or remove any components (PCBs
or Field Wiring.) while the unit is running. Doing so can
cause sudden electrical transients or imbalances that
are capable of causing damage to the module or
component in question, as well as other associated
circuit boards. Always turn off ANY Transmitter to
Instrument circuits (at the instrument or External DC
Power Source) before changing or adding any com-
ponents.
CAUTION
Place any related critical processes under manual or
auxiliary control prior to shutting down or performing
any of the steps discussed herein.
To install the Loop Powered Indicator (LPI) option into a Model 2808 Transmitter, follow
steps 1 through 9 below. To remove the LPI see step 10.
Note: The LPI is loop powered and
may only be used with Model 2808’s configured in the “4-20mA” mode.
1.
Remove the appropriate Cover Assembly from the instrument (see Figure 2-1). The
Cover Assembly is factory installed “hand tight,” i.e., there is no torque required.
Содержание Bristol 2808 Series
Страница 12: ...BLANK PAGE ...
Страница 14: ...1 2 Introduction 2808B Figure 1 1 2808 XXB Transmitters ...
Страница 16: ...BLANK PAGE ...
Страница 23: ...2808B GP Transmitters 1A 7 Figure 1A 5 Overall Dimensions Model 2808 15B With Neck Type Mounting Bracket ...
Страница 24: ...1A 8 GP Transmitters 2808B Figure 1A 6 Overall Dimensions Model 2808 16B With Neck Type Mounting Bracket ...
Страница 30: ...BLANK PAGE ...
Страница 32: ...1B 2 DP Transmitters 2808B Figure 1B 1 Model 2808 35B Figure 1B 2 Transmitter Assemblies ...
Страница 37: ...2808B DP Transmitters 1B 7 Figure 1B 4C Dimensions for Model 2808 35B Transmitter With Manifold Adapter Option ...
Страница 40: ...1B 10 DP Transmitters 2808B Figure 1B 6 Transmitter Housing Rotation Diagram ...
Страница 54: ...BLANK PAGE ...
Страница 57: ...2808B Calibration 3 3 Figure 3 2 Calibration Adjustments ...
Страница 60: ...3 6 Calibration 2808B TABLE 3 A COARSE ZERO SWITCH SETTINGS Elevation ...
Страница 61: ...2808B Calibration 3 7 TABLE 3 B COARSE ZERO SWITCH SETTINGS Suppression ...
Страница 66: ...BLANK PAGE ...
Страница 68: ...4 2 Service 2808B Figure 4 1 Using Internal TEST Terminals to Measure Current ...
Страница 72: ...BLANK PAGE ...
Страница 74: ...BLANK PAGE ...
Страница 81: ...Appendix B Page 5 Loop Powered Indicator Figure 2 2 Loop Powered Indicator Installation Diagram ...
Страница 90: ...BLANK PAGE ...
Страница 92: ...BLANK PAGE ...
Страница 100: ...BLANK PAGE ...
Страница 127: ...ESDS Manual S14006 4 15 92 CARE AND HANDLING OF PC BOARDS AND ESD SENSITIVE COMPONENTS BRISTOL BABCOCK ...
Страница 128: ...BLANK PAGE ...
Страница 139: ...BLANK PAGE ...