52
7.8 Lance replacing
A
DANGER
D10.0005
Mortal danger from electric shock
Exposed power contacts - High Voltage
−
An electric shock can cause injuries in the form of burns and cardiac
arrhythmia, respiratory arrest or cardiac arrest.
•
Set the mains switch of the analyzer to position 0 and pull the
mains plug of the mains socket.
CAUTION
C12.0076
Scalding/burns
Hot furnace / combustion tube / analyzer parts
−
Parts of the analyzer can be very hot.
•
Wait before start working until the analyzer is cooled down.
•
Switch off the analyzer and disconnect the power plug.
•
Remove the cover panel (1) of the cabinet. Unscrew 4 screws (2) on
top.
•
Remove the tubes (3), (4) and (5) from the upper furnace closure
assembly (6/7).
•
Remove the 4 screws (8) which hold the assembly (6/7) on the
furnace.
•
Lift the assembly (6/7) in straight vertical direction.
•
Unscrew the four screws and nuts which hold the parts (6) and (7)
together.
•
Replace the lance (11)
•
Replace the O-ring (10)
•
Replace the O-ring (9)
•
Do reassembling in reverse order