16
5
Installation
5.1 Setting check up
CAUTION
•
V0092
Device falling down
Incorrect erection or inadequate working space
– Due to its weight, the device can cause injuries if it falls down.
•
Only operate the device on a sufficiently large, strong, non-slip and
stable working area.
•
Ensure that all feet of the device are standing securely.
WARNING
W2.0021
Fire hazard / Risk of burns
Hot parts (crucibles, reagents,..) can fall down
•
Ignition of tables, floors, or any other surface the hot part falls on
•
Ignition of clothes and any other material
•
Set up the analyser in a flame-retardant environment. Pay special
attention to the table, the floor and any other surface being in the near
of the analyser
•
Make sure that the surrounding space of the analyser stays clear of
any flammable material
•
Always wear suitable clothing
•
Keep the work environment clear of all materials that could catch fire
•
Do not store any items on top of the analyzer
DANGER
D2.0005
Mortal danger from electric shock
Exposed power contacts - High Voltage
−
An electric shock can cause injuries in the form of burns and cardiac
arrhythmia, respiratory arrest or cardiac arrest.
•
Set the mains switch of the analyser to position 0 and pull the
mains plug of the mains socket.
NOTICE
Installation of the machine
–
It must be possible to disconnet the machine from the mains at any time.
•
Install the machine such that the connection for the mains cable is
easily accessible.
Since the analyser weighs about 90 kg it should be placed on a suitably stable and flame-
retardant surface. Due to the balance, the platform should be as free of vibration as possible.
The loader is attached to the right side of the analyser. The size of a desktop (approx. 200 cm
80 cm for the analyzer with 130-crucibles autoloader) is sufficient for the entire set-up.
Below is an example of installation:
Содержание CS-580A
Страница 1: ...1 Manual Carbon Sulfur Analyzer CS 580A HELIOS REV0004 04 19 Translation ...
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...
Страница 56: ...56 8 2 Gas flow system ...
Страница 63: ...63 9 3 Bottom outside view Part Nr Description 11035 Cooling fan 11492 Oxygen pressure regulator ...
Страница 67: ...67 9 6 TIC module 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 ...
Страница 70: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...