Maintenance
71
16
Nuts for furnace housing
17
Upper furnace lock
•
Remove the furnace cover by just loosening the nuts (16).
•
Unscrew the knurled nuts (1) and washers (2).
•
Unscrew the wing nuts (3).
•
Unscrew the nuts (4) and detach the tubes (5).
•
Remove the furnace cleaning system, by pulling the mounting (6).
•
Unscrew the nut (7) and the detach tube (8).
•
Unscrew the wing nuts (9) and pull down lower furnace lock (10).
•
Pull-off the lower O-ring (12) from the combustion tube, remove the combustion tube (14)
by pulling it up; remove the upper O-ring (11).
•
Apply a thin layer of grease on the inner surface of the new O-rings (11) and (12), before
mounting them. Apply a thin layer of grease on the outer surface of the combustion tube,
where the new O-rings will be placed.
•
Reinstall in reverse order.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...
Страница 5: ...5 ...
Страница 63: ...Maintenance 63 Fig 35 Induction furnace Cone ...
Страница 80: ...Function description 80 Fig 51 Gas flow system ...
Страница 91: ...Miscellaneous 91 8 1 6 Furnace Fig 60 Furnace 71031 71010 ...
Страница 98: ...Miscellaneous 98 8 1 12 Tic module Fig 66 TIC module 35543 9008 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 ...