Function description
84
This is important for analysis in the very low ppm range. The crucibles are inserted into the
furnace tube and they remain preheated in the tube until needed. Each time a crucible is
needed, a new one is inserted into the tube, and a preheated crucible falls out the other end of
the furnace tube. The recommended pre-heating temperature is between 1250°C and 1350°C.
For preheating crucibles the boat stop is removed.