Maintenance
67
A:
The cock (1) is rotated by 180°, so that the O-ring (2) loosens.
B:
The dust trap (3) is raised as far as it will go.
C:
Then it is swung to the side and
D:
Detached downwards at an angle.
•
Clean the dust trap (3).
•
Reinstall in reverse order after cleaning.
Fast filter cleaning
–
once every 200 analyses when using tungsten accelerator
–
once every 100 analyses when using tungsten and Iron
Fig. 39: Dust trap: fast cleaning
•
Clean the dust using the brush (1) delivered with the analyzer.
•
Rotate in only one direction.
•
Clean the upper end of the filter housing (2).
NOTICE
Grease only the lower end of the filter housing (3) and the lower O-ring (4). The upper end of
the filter housing (2) and the O-ring of the upper sealing mechanism should remain clean and
absolutely free of grease.
Thorough cleaning
–
once every 1000 analyses when using tungsten accelerator
–
once every 500 analyses when using tungsten and Iron
Fig. 40: Dust trap: disassembling
•
Remove the metal filter out of the filter housing.
•
Perform a preliminary cleaning, by using the brush.
•
Clean the metal filter in the ultrasonic cleaner.
•
Dry and, if necessary for assembling, lubricate the 0-ring.
•
Clean the upper end of the filter housing (2) from any grease.
NOTICE
When reinstalling the filter in the filter housing, the O-rings must be correctly installed otherwise
the gas flow will be completely blocked.
Outer O-ring on top, inner O-ring to the bottom.
B 6 - 3 - 3
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...
Страница 5: ...5 ...
Страница 63: ...Maintenance 63 Fig 35 Induction furnace Cone ...
Страница 80: ...Function description 80 Fig 51 Gas flow system ...
Страница 91: ...Miscellaneous 91 8 1 6 Furnace Fig 60 Furnace 71031 71010 ...
Страница 98: ...Miscellaneous 98 8 1 12 Tic module Fig 66 TIC module 35543 9008 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 ...
Страница 110: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...