Maintenance
60
1
Knurled nut
2
Washers
3
Wing nuts
4
Nut for gas inlet tube
5
Gas inlet tube
6
Mounting
7
Nut for gas outlet tube
8
Gas outlet tube
9
Wing nuts
10
Lower furnace lock
11
Upper o-ring for combustion tube
12
Lower o-ring for combustion tube
13
O-ring for lower furnace lock
14
Combustion tube
15
Induction coil
16
Nuts for furnace housing
17
Upper furnace lock
6.6.1 Replacing the o-rings 11 and 12 for combustion tube:
•
Remove the furnace housing by just loosening the nuts (16).
•
Open the furnace.
•
Unscrew the knurled nuts (1) and washers (2).
•
Unscrew the wing nuts (3).
•
Unscrew the nuts (4) and detach tubes (5).
•
Remove the furnace cleaning system, by pulling up the mounting (6).
•
Unscrew the nut (7) and detach the tube (8).
•
Unscrew the wing nuts (9) and pull down lower furnace lock (10).
•
Now the O-rings (11) and/or (12) can be removed and replaced. Apply a thin layer of
grease on the inner surface of the new O-rings, before mounting them. Apply a thin layer
of grease on the outer surface of the combustion tube, where the new O-rings will be
placed.
•
Reinstall in reverse order.
6.6.2 Replacing the o-ring 13 for lower furnace lock
•
Remove the furnace cover by just loosening the nuts (16)
•
Unscrew the nut (7) and detach tube (8)
•
Unscrew the wing nuts (9) and pull down lower furnace lock (10)
•
Remove the O-ring (13) with a screwdriver; insert a new one without greasing it.
•
Reinstall in reverse order.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...
Страница 5: ...5 ...
Страница 63: ...Maintenance 63 Fig 35 Induction furnace Cone ...
Страница 80: ...Function description 80 Fig 51 Gas flow system ...
Страница 91: ...Miscellaneous 91 8 1 6 Furnace Fig 60 Furnace 71031 71010 ...
Страница 98: ...Miscellaneous 98 8 1 12 Tic module Fig 66 TIC module 35543 9008 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 ...
Страница 110: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...