Miscellaneous
103
8.3 Pre-installation guide
WARNING
W15.0021
Fire hazard / Risk of burns
Hot parts (crucibles, reagents,...) can fall down
Ignition of tables, floors, or any other surface the hot part falls on
Ignition of clothes and any other material
•
Set up the analyser in a flame retardant environment. Pay special
attention to the table, the floor and any other surface being in the near
of the analyzer
•
Always wear suitable clothing
•
Keep the work environment clear of all materials that could catch fire
Following requirements apply, when installing the analyzer:
Carrier gas
Oxygen 99.5% pure; 2 - 4bar (30 - 60psi)
Compressed air
4
– 6 bar (60 - 90psi)
Mains power supply :
Analyzer:
230VAC ±10%, 50/60Hz; 16A fuse
(CEE-Plug 230V, 16A)
Resistance Furnace:
230 VAC ±10%, 50/60 Hz; 20A fuse
CEE-Plug 230V, 32A
Analyzer dimension
560 x 780 x 600 mm (WHD)
Furnace dimension
340 x 525 x 600 mm (WHD)
Analyzer weight
approx. 140 kg.
Furnace weight
approx. 38 kg.
–
It is important to install the instrument on a stable place.
–
The balance should rest on a vibration free support.
Содержание CS-2000
Страница 2: ...2 Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...
Страница 5: ...5 ...
Страница 63: ...Maintenance 63 Fig 35 Induction furnace Cone ...
Страница 80: ...Function description 80 Fig 51 Gas flow system ...
Страница 91: ...Miscellaneous 91 8 1 6 Furnace Fig 60 Furnace 71031 71010 ...
Страница 98: ...Miscellaneous 98 8 1 12 Tic module Fig 66 TIC module 35543 9008 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 ...
Страница 110: ...Copyright Copyright by Eltra GmbH Retsch Allee 1 5 42781 Haan Germany ...