How to replace components
on the generator print
© Elma GmbH & Co KG
SM/Transsonic_LC/1104/D
79
The capacitor in the output circle (2kV) can be destroyed due to
a fault in the transducer system or through wrong operation
(cleaning items directly on the tank floor). The ultrasonic
frequency is undefined around 50kHz.
22nF / 2KV: 200 000 0079
10nF / 2 KV: 200 000 5425
There can occur a short circuit in the coil of the secondary
output transformer, or the coil can be overheated / scorched
due to a fault in the transducer system or through wrong
operation (cleaning items directly on the tank floor). The fuse on
the generator print is intact if the fault is limited to the output
transducer.
Check the transducer system (the transducer disks might be
torn).
The oscilloscope can detect this fault by measuring the collector
current (same measuring points as for adjustment see
section
16.5.
). The graph remains flat (
see section 16.2. Illustration
16.2.4.
).
Capacitor
Elma Order No.
Output transformer
1. Open the screw connection of the output transformer.
How to proceed
2. Cut the silicone fixation of the coil with the core.
3. Remove the upper U ferrite core with its support.
4. Unsolder the secondary coil (coil to the output side).
5. Replace the coil.
6. Solder the connections of the new coil.
Observe polarization: The output connection of the coil
(outer connection) must be soldered to the contact on the
generator print (to output).
7. Put the U ferrite core on the two coils.
Make sure that the washers between the upper and the
lower U core are inserted.
8. Screw the upper U ferrite core with its suport.
Do not tighten the screws to much to prevent the ferrite core
from tearing.
9. Secure the coil with silicone.
Ensure that all cables are free so that no short circuit can
occur.
Secondary coil 115V 200 000 7211
Secondary coil 230V 200 000 7205
Elma Order No.