How to replace components
on the generator print
© Elma GmbH & Co KG
SM/Transsonic_LC/1104/D
77
100 Ohm / 0,5W: 200 000 5005
47 kOhm / 1W: 200 000 0446
The capacitor in the output circle (2kV) can be destroyed due to
a fault in the transducer system or through wrong operation
(cleaning items directly on the tank floor). The ultrasonic
frequency is undefined around 50kHz.
200 000 0042
There can occur a short circuit in the coil of the secondary
output transformer, or the coil can be overheated / scorched
due to a fault in the transducer system or through wrong
operation (cleaning items directly on the tank floor). The fuse on
the generator print is intact if the fault is limited to the output
transducer.
Check the transducer system (the transducer disks might be
torn).
The oscilloscope can detect this fault by measuring the collector
current (same measuring points as for adjustment see
section
16.4.
). The graph remains flat (
see section 16.2. Illustration
16.2.4.
).
115V
: 200 000 1460 /
230V
: 200 000 1459
Illustration 15.2.1 Generator print with relevant components
Transistors
Mains rectifyer diodes 1N4007 (4 pieces)
Varistor
Output transformer
Output capacitors
Capacitor
Elma Order No.
Output transformer
Elma Order No.
A
C
D
E
B
A
B
C
D
E