Trouble shooting / List of malfunctions
© Elma GmbH & Co KG
SM/Transsonic_LC/1104/D
11
6.1 T310(H)
Malfunction
1. Check:
Heating
switch
- if lighted
2. Check:
Fuse
printed
board
Possible causes
Remedy
Check the unit for possible damages and traces of liquid inside (e.g. leak in the
tank, liquid/humidity entering from outside)
No ultrasound
(Heating OK)
OK
OK
•
interference filter
coil
•
timer
•
replace component
•
replace complete
interference filter
No ultrasound /
weak ultrasound
(Heating OK)
OK
OK
•
output transformer
•
output capacitor
6.8nF
•
scorching in
transducer system
•
replace component
•
replace complete
printed board
•
open/check
transducer system
No ultrasound
No heating
OFF
faulty
•
faulty printed board
(BZW diode;
transistors; diodes
mains and HF)
•
replace component
•
replace complete
printed board
•
disconnect heating
cable,
check unit again (new
fuses)
if short circuit does not
occur again, the
heating is faulty
•
replace faulty heating
No ultrasound
No heating
OFF
OK
•
interference filter
•
mains cable
•
replace complete
printed board
•
replace mains cable