Putting into operation / Trial run
© Elma GmbH & Co KG
SM/Transsonic_LC/1104/D
103
17
Putting into operation / Trial run
For checking and trial operation of the unit after repair please
observe the instructions given in this section. We recommend a
trial run over several hours (approx. 3 – 4 h) after repair.
Place the unit on a dry and stable workplace. Ensure that the
workplace is sufficiently ventilated!
CAUTION
Risk of electrocution due to entering humidity!
Keep the unit dry.
The unit inside is splash-proof. To prevent electrical accidents
and damage keep the workplace and the unit housing dry.
Ambient conditions
•
Allowed ambient temperature during operation:
+5°C to +40°C
•
Allowed relative humidity of air during operation: 80%
•
No condensation allowed
The unit must be connected to a shockproof socket only.
Mains connection
17.1
How to fill cleaning liquid
Fill the cleaning tank of the unit with a sufficient quantity of
liquid before switch-on.
Observe filling level
Dry running of the transducer tank can damage the unit.
Ensure that the cleaning tank is correctly filled during operation
(mark on the tank wall).
NOTE
Make sure that the selected cleaning chemical is perfectly
suitable for the use in an ultrasonic bath and follow the
instructions on dosage and material compatibility
.
In general, flammable products are not allowed. Observe the
safety warnings in
section 17.2.
The sefety warnings include instructions on limitations
concerning cleaning media in
section 17.2.
Fresh cleaning liquid is saturated with air which impairs the
cleaning effect of the ultrasonic activity. Sounding the liquid
over a period of several minutes before the actual cleaning
process is started will eliminate the microscopic air bubbles
from the liquid.
Suitable cleaning
media
Prohibited cleaning
media
How to degas the
liquid