How to replace the interference filter
© Elma GmbH & Co KG
SM/Transsonic_LC/1104/D
59
13
How to replace the interference filter
A fault on the interference filter (pcb) is very rare. In most
cases, the cause of a malfunction is a blown fuse or fuses.
As a rule, faulty fuses are caused by a short circuit in a heating
element or by humidity inside the unit. The units T310, T420,
LC20 and LC30 do not have a separate interference filter, in
these units the breakdown can be caused by a fault on the
generator.
On older units without fuse it is possible that a short circuit
current has interrupted the interference filter coil.
13.1
Interference filter on T310,T420, LC20 and LC30
The interference filter is integrated on the generator pcb. A
breakdown of the fuse can be caused by a short circuit in the
heating element or the final stage of the generator.
In order to find the cause we recommend to interrupt the cables
of the heating and to test the unit again with new fuses.
Illustration 13.1.1 Generator pcb complete with interference filter
(section marked yellow)
Supply cable interference filter >> timer input.
Polarization irrelevant. On units with heating the cable for the
heating is connected to the twin contacts.
Supply cable timer output >> generator.
Polarization irrelevant.
Mains supply phase (brown).
Interference filter
section on the
generator pcb
A
B
D
E
F
G
H
C
A-B
C-D
E