Operating faults
© Elma Schmidbauer GmbH
BA_Elmasolvex_VA_EN_Vers.05.2018
97
Warning!
After pressing the
button with the warning
symbol:
heater was switched off
during the drying as
liquid was detected.
lower fill level sensor (or this
sensor is defective) during
the drying, the heater is
switched off and the warning
message is displayed.
•
Filling level sensor
Empty level
(installed under the sieve insert)
(
Fig.9.1.4.1.F
) soiled? Clean with
suitable cloth if necessary.
•
Fill level sensor
Empty level
(
Fig.9.1.4.1.F
) damaged (scratches,
cracks)?
•
Retrofit anti-reflection sleeve for
Empty level
fill level sensor.
•
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
299
Warning!
Contact the service.
Change of the rotation
seal is necessary!
See operating
instructions.
Check the following possibilities:
•
Take machine out of operation
immediately.
•
Contact Service
Safe operation of the cleaning machine is
no longer guaranteed. The required
service must strictly be performed to
continue operating the cleaning machine.
The manufacturer shall not accept any
liability for injuries to persons and property
damage resulting from further operation.