Operating faults
86
BA_Elmasolvex_VA_EN_Vers.05.2018
© Elma Schmidbauer GmbH
•
Temperature of the medium in the
media tank, which the indicated partial
step in the error refers to, already too
high? (see temperature table for
Warning 211
).
Remedy: use cooled down medium for
the indicated process step.
•
Gas content in cleaning medium too
high? Switch off the ultrasound and
basket oscillation temporarily to degas
the cleaning medium more slowly.
•
For some programs: Check deflection
and frequency rate oscillation
parameters.
•
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
115-146
Warning
A fan is not functioning
correctly.
Program is not ended
properly. The cleaning
unit cannot be restarted
while the fault is
present.
A fan is blocked or defective. Check the following possibilities:
•
Check fans for foreign bodies and
remove if necessary.
•
Switch off cleaning machine and
restart: Contact Service if the fault
persists.
147-178
Warning
Working level filling
level not correct.
Fill level in cleaning chamber
not optimal (or fluctuates
from time to time).
Check the following possibilities:
•
Cleaning basket inserted correctly?
•
Sieve insert (
Fig.9.1.4.1.D
) in the
cleaning chamber inserted correctly?
•
Fill level in the media tank is too low
(and air is drawn in as a result)?
•
Working level fill level sensor and
overfill level fill level sensor
(
Fig.9.1.4.1 A/B
) soiled?
Clean with suitable cloth if necessary.
•
Working level fill level sensor and
overfill level fill level sensor
(
Fig.9.1.4.1 A/B
) damaged (scratches,
cracks)?
•
Quick couplings (
Fig. 3.13.1. B/C
)
engaged?
•
Flat seal in quick coupling (
Fig.
4.4.1.4
) damaged?
•
Gas content in cleaning medium too
high? Switch off the ultrasound and
basket oscillation temporarily to degas
the cleaning medium.
•
Switch off cleaning machine and
restart: Contact Service if the fault
persists.