Extended safety test
An extended safety test must be performed after every 500
cleaning cycles or at the latest every 3 months.
A corresponding message is shown on the display. This must
be confirmed with to start the safety test.
In each case the execution of the safety test can be postponed
three times with ; then the safety test must be started
(
Fig. 10.3.1
). Another start of the cleaning machine is not
possible again until after the extended safety test has been
performed.
Fig. 10.3.1 Display
Start safety test
For this extended safety test, the function of the
overfilling level
filling level sensor is tested by suctioning medium.
In the case of possible malfunction of the overfilling level
sensor, medium can escape. This should be collected by the
collecting container.
It is required here that the exhaust air tube included in the
scope of delivery is connected to the ventilation connection of
the cleaning chamber (
Fig. 4.3.1.A
).
The end of the exhaust
air tube must be immersed in a collecting container (>= 1.5
litres) during this test.
Before performing the safety test, ensure that media tank #4 is
filled between the Min and Max markings.