Before the initial commissioning
© Elma Schmidbauer GmbH
BA_Elmasolvex_VA_EN_Vers.05.2018
27
The filling level of the operating materials must always be
between the markings (
Fig. 4.4.3.D
) to guarantee proper
operation of the cleaning machine and an optimal cleaning
result.
Too low a filling level (below the
Min
marking) causes errors in
the cleaning program with unsatisfactory results.
Overfilling (above the
Max
marking) can result in liquid
discharging at the case outlet at the bottom and possibly
destroying the activated carbon unit.
The volume above the
Max
marking is also used as reserve for
possible liquid entrainments.
5. Check the shut-off valves of the quick couplings on the
media tank and the media tubes for correct function
(
Chapter 4.4.1
).
6. Replace the tank in the cleaning machine.
7. Reconnect the tube couplings to the media tanks. The tube
couplings must audibly engage in the shut-off valves.
Also visually inspect the secure seating of the tube
couplings.
Attention!
After the engagement, the push buttons on the
tube couplings (B and C) must be completely disengaged.
Check by shaking and pulling on the tube coupling whether
it is correctly connected.
8. Proceed in the same way with the media tanks #2 - #3 - #4
for the rinsing media.
Fig. 4.4.2 Media tank cover opened
ATTENTION
A
#1
#2
A
C
B